To assess the agreement between a new automatic device (FS-20D) using a cuff-oscillometric method to measure arterial blood pressure (BP) in the fingers and a standard mercury sphygmomanometer.
The blood pressure measurements were taken in a sequential order, in a sample of both normotensive subjects (n. 57) and slight to moderate hypertensive patients (n. 28) without vascular complications.
The mean sphygmomanometer-monitor difference was 0.52 +/- 4.57 mmHg for systolic and 0.25 +/- 4.41 mmHg for diastolic values; the agreement limits were: SBP -8.6 divided by 9.6 mmHg, 95% CI: -0.5 divided by 1.5; DBP: -8.6 divided by 9.1 mmHg, 95% CI: -0.7 divided by 1.2. The grade of agreement between the monitor and the sphygmomanometer was "A" (British Hypertension Society) for both systolic and diastolic values (difference of readings < 5 mmHg: 82%; < 10 mmHg: 97% for systolic blood pressure, 98% of diastolic blood pressure).
The monitor was proved to be reliable with a good level of precision and accuracy. The FS-20D monitor may be used in self-monitoring of blood pressure of patients with slight to moderate hypertension.
评估一种采用袖带示波法测量手指动脉血压的新型自动设备(FS - 20D)与标准汞柱血压计之间的一致性。
对无血管并发症的正常血压受试者(57例)和轻度至中度高血压患者(28例)样本依次进行血压测量。
收缩压方面,血压计与监测仪的平均差值为0.52±4.57 mmHg,舒张压方面为0.25±4.41 mmHg;一致性界限为:收缩压-8.6至9.6 mmHg,95%置信区间:-0.5至1.5;舒张压:-8.6至9.1 mmHg,95%置信区间:-0.7至1.2。监测仪与血压计在收缩压和舒张压值方面的一致性等级均为“A”(英国高血压学会)(读数差值<5 mmHg:82%;<10 mmHg:收缩压为97%,舒张压为98%)。
该监测仪被证明具有良好的精密度和准确性,是可靠的。FS - 20D监测仪可用于轻度至中度高血压患者的血压自我监测。