Diessel E, Fuerst T, Njeh C F, Hans D, Cheng S, Genant H K
Osteoporosis and Arthritis Research Group, University of California at San Francisco 94117, USA.
Br J Radiol. 2000 Jan;73(865):23-30. doi: 10.1259/bjr.73.865.10721316.
The purpose of this study was to evaluate a new imaging ultrasound scanner for the heel, the DTU-one (Osteometer MediTech, Denmark), by comparing quantitative ultrasound (QUS) results with bone mineral density (BMD) of the heel and femur from dual X-ray absorptiometry (DXA), and by comparing the DTU-one with another QUS device, the UBA 575+. The regions of interest in the DXA heel scan were matched with the regions evaluated by the two QUS devices. 134 healthy and 16 osteoporotic women aged 30-84 years old were enrolled in the study. In vivo short-term precision of the DTU-one for broadband ultrasound attenuation (BUA) and speed of sound (SOS) was 2.9% and 0.1%, respectively, and long-term precision was 3.8% and 0.2%, respectively. Highest correlations (r) between QUS and BMD measurements were achieved when comparing DTU-one results with BMD in matched regions of the DXA heel scan. Correlation coefficients (r) were 0.81 for BUA and SOS. Highest correlations with the UBA 575+ were 0.68 and 0.72, respectively. The comparison of BMD in different femoral sites with BUA and SOS (DTU-one) varied from 0.62 to 0.69 when including the entire study population. The correlation between BMD values within different sites of the femur tended to be higher (from r = 0.81 to 0.93). When comparing BUA with BUA and SOS with SOS on the two QUS devices, the absolute QUS values differed significantly. However, correlations were relatively high, with 0.76 for BUA and 0.82 for SOS. In conclusion, the results of the new quantitative ultrasound device, the DTU-one, are highly correlated (r = 0.8) with results obtained using the UBA 575+ and with BMD in the heel. The precision of the DTU-one is comparable to other QUS devices for BUA and is high for SOS.
本研究的目的是通过将定量超声(QUS)结果与双能X线吸收法(DXA)测量的足跟和股骨骨密度(BMD)进行比较,并将DTU-one与另一台QUS设备UBA 575+进行比较,来评估一款用于足跟的新型超声成像扫描仪DTU-one(丹麦Osteometer MediTech公司)。DXA足跟扫描中的感兴趣区域与两台QUS设备评估的区域相匹配。134名年龄在30 - 84岁的健康女性和16名骨质疏松女性参与了本研究。DTU-one对宽带超声衰减(BUA)和声速(SOS)的体内短期精密度分别为2.9%和0.1%,长期精密度分别为3.8%和0.2%。将DTU-one的结果与DXA足跟扫描匹配区域的BMD进行比较时,QUS与BMD测量之间的相关性最高(r)。BUA和SOS的相关系数(r)均为0.81。与UBA 575+的最高相关性分别为0.68和0.72。当纳入整个研究人群时,不同股骨部位的BMD与BUA和SOS(DTU-one)的比较结果在0.62至0.69之间。股骨不同部位的BMD值之间的相关性往往更高(r从0.81至0.93)。在两台QUS设备上比较BUA与BUA以及SOS与SOS时,QUS的绝对值存在显著差异。然而,相关性相对较高,BUA为0.76,SOS为0.82。总之,新型定量超声设备DTU-one的结果与使用UBA 575+获得的结果以及足跟的BMD高度相关(r = 0.8)。DTU-one对BUA的精密度与其他QUS设备相当,对SOS的精密度较高。