Kost G J, Vu H T, Inn M, DuPlantier R, Fleisher M, Kroll M H, Spinosa J C
University of California, Davis, Health System and the Point-of-Care Testing Center for Teaching and Research, USA.
Crit Care Med. 2000 Jul;28(7):2379-89. doi: 10.1097/00003246-200007000-00033.
To introduce a creatinine biosensor and a total carbon dioxide content (TCO2) method for whole-blood measurements, to evaluate the clinical performance of a new transportable analyzer that simultaneously performs these two and six other tests (Na+, K+, Cl-, glucose, urea nitrogen, and hematocrit), and to assess the potential of the new analyzer for point-of-care testing in critical care by comparing results obtained by nonlaboratory personnel and by medical technologists.
Multicenter sites compared whole-blood measurements with the transportable analyzer to plasma measurements from the same specimens with local reference instruments. One site compared whole-blood results produced by nonlaboratory personnel vs. medical technologists and evaluated day-to-day and within-day precision at the point of care.
Four medical centers in the United States. Venous and arterial specimens from 710 critically ill patients with a variety of diagnoses. Point-of-care testing in the emergency room and operating room.
The linear regression analyses at the four medical centers showed the following: creatinine (a) slope, 0.91 to 1.22, (b) y intercept, -0.07 to 0.15 mg/dL, and (c) r2, 0.77 to 1.00; and TCO2: (a) slope, 0.64 to 1.00, (b) y intercept, 1.36 to 9.6 mmol/L, and (c) r2, 0.52 to 0.72 (yi, whole-blood analyses; xi, plasma reference measurements). Bland-Altman plots also were used to assess multicenter creatinine and TCO2 results. Of the other analytes, K+, glucose, and urea nitrogen had the highest r2-values. For the eight chemistry profile tests performed at the point of care (yi, nonlaboratory personnel results; xi, medical technologist results), the average value of r2 was 0.96 (SD 0.08) in the operating room and 0.96 (SD 0.06) in the emergency room, and mean paired differences (yi - xi) were not statistically or clinically significant. Precision was acceptable.
The performance of the creatinine biosensor and the TCO2 method was acceptable for whole-blood samples. Comparisons of whole-blood results from the transportable analyzer and plasma results from the local reference instruments revealed analyte biases that may be attributed to differences between direct whole-blood analyses and indirect-diluted plasma measurements and other factors. Performance of nonlaboratory personnel and medical technologists was equivalent for point-of-care testing in critical care settings. The whole-blood analyzer should be useful when patient care demands immediate results.
介绍一种用于全血检测的肌酐生物传感器和总二氧化碳含量(TCO2)检测方法,评估一种新型便携式分析仪的临床性能,该分析仪可同时进行这两项检测以及其他六项检测(Na+、K+、Cl-、葡萄糖、尿素氮和血细胞比容),并通过比较非实验室人员和医学技术人员获得的结果,评估该新型分析仪在重症监护现场即时检测中的潜力。
多中心将便携式分析仪的全血检测结果与同一标本用当地参考仪器检测的血浆结果进行比较。一个中心比较了非实验室人员和医学技术人员得出的全血检测结果,并评估了现场即时检测的日间和日内精密度。
美国的四个医疗中心。采集了710例患有各种疾病的重症患者的静脉和动脉标本。在急诊室和手术室进行现场即时检测。
四个医疗中心的线性回归分析显示如下:肌酐(a)斜率为0.91至1.22,(b)y轴截距为-0.07至0.15mg/dL,(c)r2为0.77至1.00;TCO2:(a)斜率为0.64至1.00,(b)y轴截距为1.36至9.6mmol/L,(c)r2为0.52至0.72(yi为全血分析结果;xi为血浆参考测量结果)。Bland-Altman图也用于评估多中心肌酐和TCO2结果。在其他分析物中,K+、葡萄糖和尿素氮的r2值最高。对于在现场即时检测中进行的八项化学指标检测(yi为非实验室人员结果;xi为医学技术人员结果),手术室中r2的平均值为0.96(标准差0.08),急诊室中为0.96(标准差0.06),平均配对差异(yi - xi)在统计学或临床上均无显著意义。精密度可接受。
肌酐生物传感器和TCO2检测方法对于全血样本的性能是可接受的。便携式分析仪的全血检测结果与当地参考仪器的血浆检测结果比较显示,分析物偏差可能归因于直接全血分析与间接稀释血浆测量之间的差异以及其他因素。在重症监护现场即时检测中,非实验室人员和医学技术人员的表现相当。当患者护理需要即时结果时,全血分析仪应会很有用。