Suppr超能文献

注射异硫蓝后术中脉搏血氧饱和度读数下降。

Intraoperative decrease in pulse oximeter readings following injection of isosulfan blue.

作者信息

Hoskin R W, Granger R

机构信息

Department of Anaesthesia, St Mary's Hospital, New Westminster, BC, Canada.

出版信息

Can J Anaesth. 2001 Jan;48(1):38-40. doi: 10.1007/BF03019812.

Abstract

PURPOSE

Sentinel lymph node mapping, involving injection of isosulfan blue dye around a tumour, is beginning to be used in patients with carcinoma of the breast. Absorption of the dye into the circulation may interfere with pulse oximetry, causing falsely low readings. This report describes changes in pulse oximeter readings following injection of isosulfan blue for sentinel lymph node mapping in a patient with carcinoma of the breast.

CLINICAL FEATURES

An 83-yr-old female patient underwent sentinel node biopsy of the axilla followed by partial mastectomy for carcinoma of the left breast. Isosulfan blue was injected in the area of the tumour in the left breast. The SpO2 began to decrease 15 min after dye injection, reaching a nadir of 89-90% 30 min after injection. Arterial blood gas analysis showed normal arterial partial pressure of oxygen. Pulse oximeter readings did not return to normal until more than six hours after dye injection.

CONCLUSION

Review of the literature reveals a small number of case reports of similar occurrences of low pulse oximeter readings following injection of isosulfan blue or patent blue dye for lymphatic mapping. Data from these reports and the case described here suggest that the latency, magnitude and duration of effect on pulse oximeter readings following injection of these dyes is highly variable. It is important to rule out other causes of low pulse oximeter readings when this effect occurs; normal oxygenation can be verified with arterial blood gas analysis. Co-oximetry can be done to rule out methemoglobinemia as a cause of decreased SpO2.

摘要

目的

前哨淋巴结定位,即将异硫蓝染料注射到肿瘤周围,已开始应用于乳腺癌患者。染料吸收入循环系统可能干扰脉搏血氧饱和度测定,导致读数假性降低。本报告描述了一名乳腺癌患者在注射异硫蓝进行前哨淋巴结定位后脉搏血氧饱和度读数的变化。

临床特征

一名83岁女性患者接受了腋窝前哨淋巴结活检,随后因左乳腺癌接受了部分乳房切除术。异硫蓝注射于左乳房肿瘤区域。染料注射后15分钟,脉搏血氧饱和度开始下降,注射后30分钟降至最低点,为89%~90%。动脉血气分析显示动脉血氧分压正常。直到染料注射后六个多小时,脉搏血氧饱和度读数才恢复正常。

结论

文献回顾显示,有少数病例报告了在注射异硫蓝或专利蓝染料进行淋巴管造影后出现类似的低脉搏血氧饱和度读数情况。这些报告及本文所述病例的数据表明,注射这些染料后对脉搏血氧饱和度读数影响的潜伏期、程度和持续时间差异很大。当出现这种影响时,排除其他导致脉搏血氧饱和度读数低的原因很重要;可通过动脉血气分析来证实氧合正常。可进行一氧化碳血氧定量法以排除高铁血红蛋白血症作为脉搏血氧饱和度降低的原因。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验