Krisher K K, Gibb P, Corbett S, Church D
Children's Medical Center of Dallas, Dallas, Texas 75235, USA.
J Clin Microbiol. 2001 Aug;39(8):2880-3. doi: 10.1128/JCM.39.8.2880-2883.2001.
The performance of the BacT/Alert PF (Organon-Teknika Corp., Durham, N.C.), a new nonvented pediatric FAN blood culture bottle, was compared to that of the original pediatric bottle, the Pedi-BacT, with matched aerobic cultures obtained from two separate facilities. A total of 244 clinically significant isolates were recovered from 4,015 compliant pairs. Among the positive cultures, 170 (70%) isolates were detected in both the BacT/Alert PF and the Pedi-BacT bottles, while 47 (19%) isolates were recovered in the BacT/Alert PF bottle only and 27 (11%) isolates were recovered in the Pedi-BacT bottle only. Although isolation of specific microorganisms was comparable for the two bottles, the total number of organisms recovered by the BacT/Alert PF was greater than that by the Pedi-BacT (P = 0.0272). In addition, more organisms were recovered by the BacT/Alert PF bottle from the blood of patients receiving antimicrobial therapy (P = 0.0180). Overall time to detection was similar for the two bottles; however, a significantly decreased mean time to detection was recorded for yeast from the BacT/Alert PF bottle (22.9 h; P = 0.0001) and staphylococci from the Pedi-BacT bottle (22.5 h; P = 0.0056). One false-negative culture and five false-positive cultures occurred with the Pedi-BacT bottle, compared to one false-positive culture with the BacT/Alert PF bottle. The BacT/Alert PF bottle is a reliable blood culture bottle for pediatric blood culture specimens and may offer improved recovery of microbes from patients on antimicrobial therapy. The use of the nonvented bottle will both facilitate bottle processing and decrease expenditures for materials due to the elimination of the venting needles required for the original vented bottles.
将新型无通风口儿科FAN血培养瓶BacT/Alert PF(Organon-Teknika公司,北卡罗来纳州达勒姆)的性能与原来的儿科血培养瓶Pedi-BacT进行了比较,从两个不同机构获取了配对的需氧培养物。从4015对合格培养物中总共分离出244株具有临床意义的菌株。在阳性培养物中,170株(70%)菌株在BacT/Alert PF瓶和Pedi-BacT瓶中均被检测到,而47株(19%)菌株仅在BacT/Alert PF瓶中被分离出,27株(11%)菌株仅在Pedi-BacT瓶中被分离出。尽管两种瓶子对特定微生物的分离情况相当,但BacT/Alert PF回收的微生物总数多于Pedi-BacT(P = 0.0272)。此外,BacT/Alert PF瓶从接受抗菌治疗患者的血液中回收的微生物更多(P = 0.0180)。两种瓶子的总体检测时间相似;然而,BacT/Alert PF瓶中酵母的平均检测时间显著缩短(22.9小时;P = 0.0001),Pedi-BacT瓶中葡萄球菌的平均检测时间显著缩短(22.5小时;P = 0.0056)。Pedi-BacT瓶出现了1例假阴性培养物和5例假阳性培养物,而BacT/Alert PF瓶出现了1例假阳性培养物。BacT/Alert PF瓶是用于儿科血培养标本的可靠血培养瓶,对于接受抗菌治疗的患者,它可能能更好地回收微生物。使用无通风口的瓶子将便于瓶子处理,并由于无需原来通风口瓶子所需的排气针而降低材料成本。