Suppr超能文献

BacT/Alert PF儿科FAN血培养瓶与标准儿科血培养瓶Pedi-BacT的比较。

Comparison of the BacT/Alert PF pediatric FAN blood culture bottle with the standard pediatric blood culture bottle, the Pedi-BacT.

作者信息

Krisher K K, Gibb P, Corbett S, Church D

机构信息

Children's Medical Center of Dallas, Dallas, Texas 75235, USA.

出版信息

J Clin Microbiol. 2001 Aug;39(8):2880-3. doi: 10.1128/JCM.39.8.2880-2883.2001.

Abstract

The performance of the BacT/Alert PF (Organon-Teknika Corp., Durham, N.C.), a new nonvented pediatric FAN blood culture bottle, was compared to that of the original pediatric bottle, the Pedi-BacT, with matched aerobic cultures obtained from two separate facilities. A total of 244 clinically significant isolates were recovered from 4,015 compliant pairs. Among the positive cultures, 170 (70%) isolates were detected in both the BacT/Alert PF and the Pedi-BacT bottles, while 47 (19%) isolates were recovered in the BacT/Alert PF bottle only and 27 (11%) isolates were recovered in the Pedi-BacT bottle only. Although isolation of specific microorganisms was comparable for the two bottles, the total number of organisms recovered by the BacT/Alert PF was greater than that by the Pedi-BacT (P = 0.0272). In addition, more organisms were recovered by the BacT/Alert PF bottle from the blood of patients receiving antimicrobial therapy (P = 0.0180). Overall time to detection was similar for the two bottles; however, a significantly decreased mean time to detection was recorded for yeast from the BacT/Alert PF bottle (22.9 h; P = 0.0001) and staphylococci from the Pedi-BacT bottle (22.5 h; P = 0.0056). One false-negative culture and five false-positive cultures occurred with the Pedi-BacT bottle, compared to one false-positive culture with the BacT/Alert PF bottle. The BacT/Alert PF bottle is a reliable blood culture bottle for pediatric blood culture specimens and may offer improved recovery of microbes from patients on antimicrobial therapy. The use of the nonvented bottle will both facilitate bottle processing and decrease expenditures for materials due to the elimination of the venting needles required for the original vented bottles.

摘要

将新型无通风口儿科FAN血培养瓶BacT/Alert PF(Organon-Teknika公司,北卡罗来纳州达勒姆)的性能与原来的儿科血培养瓶Pedi-BacT进行了比较,从两个不同机构获取了配对的需氧培养物。从4015对合格培养物中总共分离出244株具有临床意义的菌株。在阳性培养物中,170株(70%)菌株在BacT/Alert PF瓶和Pedi-BacT瓶中均被检测到,而47株(19%)菌株仅在BacT/Alert PF瓶中被分离出,27株(11%)菌株仅在Pedi-BacT瓶中被分离出。尽管两种瓶子对特定微生物的分离情况相当,但BacT/Alert PF回收的微生物总数多于Pedi-BacT(P = 0.0272)。此外,BacT/Alert PF瓶从接受抗菌治疗患者的血液中回收的微生物更多(P = 0.0180)。两种瓶子的总体检测时间相似;然而,BacT/Alert PF瓶中酵母的平均检测时间显著缩短(22.9小时;P = 0.0001),Pedi-BacT瓶中葡萄球菌的平均检测时间显著缩短(22.5小时;P = 0.0056)。Pedi-BacT瓶出现了1例假阴性培养物和5例假阳性培养物,而BacT/Alert PF瓶出现了1例假阳性培养物。BacT/Alert PF瓶是用于儿科血培养标本的可靠血培养瓶,对于接受抗菌治疗的患者,它可能能更好地回收微生物。使用无通风口的瓶子将便于瓶子处理,并由于无需原来通风口瓶子所需的排气针而降低材料成本。

相似文献

2
Controlled comparison of original vented aerobic fan medium with new nonvented BacT/ALERT FA medium for culturing blood.
J Clin Microbiol. 2001 Jun;39(6):2098-101. doi: 10.1128/JCM.39.6.2098-2101.2001.
3
Controlled clinical comparison of the BacT/ALERT FN and the standard anaerobic SN blood culture medium.
J Clin Microbiol. 2004 Oct;42(10):4581-5. doi: 10.1128/JCM.42.10.4581-4585.2004.
5
Controlled clinical comparison of VersaTREK and BacT/ALERT blood culture systems.
J Clin Microbiol. 2007 Feb;45(2):299-302. doi: 10.1128/JCM.01697-06. Epub 2006 Nov 22.
10
Three days of incubation may be sufficient for routine blood cultures with BacT/Alert FAN blood culture bottles.
J Clin Microbiol. 2001 Jun;39(6):2079-82. doi: 10.1128/JCM.39.6.2079-2082.2001.

引用本文的文献

2
Ileus in children presenting with diarrhea and severe acute malnutrition: A chart review.
PLoS Negl Trop Dis. 2017 May 11;11(5):e0005603. doi: 10.1371/journal.pntd.0005603. eCollection 2017 May.
4
Blood culture isolates in neonatal sepsis and their sensitivity in Anand District of India.
Indian J Pediatr. 2014 Aug;81(8):785-90. doi: 10.1007/s12098-013-1314-2. Epub 2014 Jan 11.

本文引用的文献

1
Controlled comparison of original vented aerobic fan medium with new nonvented BacT/ALERT FA medium for culturing blood.
J Clin Microbiol. 2001 Jun;39(6):2098-101. doi: 10.1128/JCM.39.6.2098-2101.2001.
2
Frequency of low-level bacteremia in children from birth to fifteen years of age.
J Clin Microbiol. 2000 Jun;38(6):2181-5. doi: 10.1128/JCM.38.6.2181-2185.2000.
3
Comparison of the BacT/Alert FAN aerobic and the Difco ESP 80A aerobic bottles for pediatric blood cultures.
J Clin Microbiol. 1997 May;35(5):1166-71. doi: 10.1128/jcm.35.5.1166-1171.1997.
4
Clinical importance of increased sensitivity of BacT/Alert FAN aerobic and anaerobic blood culture bottles.
J Clin Microbiol. 1996 Sep;34(9):2180-4. doi: 10.1128/jcm.34.9.2180-2184.1996.
5
Effect of number of blood cultures and volume of blood on detection of bacteremia in children.
J Pediatr. 1996 Feb;128(2):190-5. doi: 10.1016/s0022-3476(96)70388-8.
7
Pediatric blood cultures.
Clin Lab Med. 1994 Mar;14(1):17-30.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验