Suppr超能文献

对由三维计算机化治疗计划系统执行的监测单位计算进行独立验证。

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system.

作者信息

Leszczynski K W, Dunscombe P B

机构信息

Department of Medical Physics, Northeastern Ontario Regional Cancer Centre, 41 Ramsey Lake Road, Sudbury, P3E 5J1, Canada.

出版信息

J Appl Clin Med Phys. 2000 Autumn;1(4):120-5. doi: 10.1120/jacmp.v1i4.2633.

Abstract

The checking of monitor unit calculations is recognized as a vital component of quality assurance in radiotherapy. Using straightforward but detailed computer-based verification calculations it is possible to achieve a precision of 1% when compared with a three-dimensional (3D) treatment planning system monitor unit calculation. The method is sufficiently sensitive to identify significant errors and is consistent with current recommendations on the magnitude of uncertainties in clinical dosimetry. Moreover, the approach is accurate in the sense of being highly consistent with the validated 3D treatment planning system's calculations.

摘要

监测单元计算的核查被认为是放射治疗质量保证的重要组成部分。通过使用简单但详细的基于计算机的验证计算,与三维(3D)治疗计划系统的监测单元计算相比,可实现1%的精度。该方法足够灵敏,能够识别重大误差,并且与当前关于临床剂量测定不确定度大小的建议一致。此外,从与经过验证的3D治疗计划系统的计算高度一致的意义上来说,该方法是准确的。