Responses of a direct ion storage dosimeter (DIS-1) to heavy charged particles.

作者信息

Yasuda H

机构信息

National Institute of Radiological Sciences, Anagawa 4-9-1, Inage, Chiba-shi 263-8555, Japan.

出版信息

Radiat Res. 2001 Dec;156(6):805-8. doi: 10.1667/0033-7587(2001)156[0805:roadis]2.0.co;2.

Abstract

The responses of a direct ion storage dosimeter (DIS-1) to energetic heavy charged particles were examined using (4)He, (12)C, (40)Ar and (56)Fe ion beams at the HIMAC at the National Institute of Radiological Sciences. The efficiency of the DIS-1 on the basis of absorbed dose was almost unity for the helium and carbon ions and was slightly decreased for the argon and iron ions. The linearity in the dose response and the angular independence for these heavy ions were fairly good. Although further studies are necessary, these results suggest that the DIS-1 would be a suitable passive dosimeter for measurements of absorbed dose in a field dominated by heavy charged particles such as the space environment.