用于骨水泥型髋关节和膝关节植入物的Ti-6Al-4V与聚甲基丙烯酸甲酯之间的静摩擦系数。

Static coefficient of friction between Ti-6Al-4V and PMMA for cemented hip and knee implants.

作者信息

Nuño N, Amabili M, Groppetti R, Rossi A

机构信息

Département de Génie Mécanique, Ecole de Technologie Supérieure, Université du Québec, 1100 Notre-Dame Str. W., Montréal, Québec, H3C 1K3, Canada.

出版信息

J Biomed Mater Res. 2002 Jan;59(1):191-200. doi: 10.1002/jbm.1233.

Abstract

The static coefficient of friction between Ti-6Al-4V and PMMA was determined experimentally. A microtopographic surface analysis of the Ti-6Al-4V and PMMA specimens used in the experiments was performed to characterize the surfaces. The coefficient of friction between Ti-6Al-4V and PMMA in both dry and wet conditions, using both Ringer's solution and bovine serum, was determined by the standard inclined plane test, following the ASTM 4516-91 method, and by a prototype computerized sliding friction tester. The effects of surface roughness and of contact pressure on the coefficient of friction also have been investigated. Tests were performed at 26 degrees C and at body temperature of 37 degrees C. Considering all the tests, the overall range of the mean coefficients of friction varied between 0.17 and 0.32 in dry or wet conditions. For the same surface roughness in contact, in general the coefficient of friction using Ringer's solution was slightly lower than it was in dry conditions whereas bovine serum had a very high surface tension, which significantly increased the static coefficient of friction.

摘要

通过实验测定了Ti-6Al-4V与聚甲基丙烯酸甲酯(PMMA)之间的静摩擦系数。对实验中使用的Ti-6Al-4V和PMMA试样进行了微观形貌表面分析,以表征其表面特征。按照ASTM 4516 - 91方法,通过标准斜面试验以及一台原型计算机化滑动摩擦测试仪,测定了在干燥和湿润条件下(使用林格氏溶液和牛血清)Ti-6Al-4V与PMMA之间的摩擦系数。还研究了表面粗糙度和接触压力对摩擦系数的影响。试验在26摄氏度和37摄氏度的体温下进行。综合所有试验来看,在干燥或湿润条件下,平均摩擦系数的总体范围在0.17至0.32之间。对于相同粗糙度的接触面,一般来说,使用林格氏溶液时的摩擦系数略低于干燥条件下的摩擦系数,而牛血清具有非常高的表面张力,这显著提高了静摩擦系数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验