High-performance liquid chromatography applications of optical rotation detection with compensation for scattering and absorbance at the laser wavelength.

作者信息

Driffield M, Bergström E T, Goodall D M, Klute A S, Smith D K

机构信息

Department of Chemistry, University of York, Heslington, UK.

出版信息

J Chromatogr A. 2001 Dec 21;939(1-2):41-8. doi: 10.1016/s0021-9673(01)01336-x.

Abstract

Use of instrumentation developed to enable simultaneous monitoring of optical rotation (OR) and transmittance allows OR measurements to be made in the presence of high levels of absorbance, scattering or other effects that change the intensity of the plane-polarised light at the photodiode detector. This extends the application of OR detection to areas where it was previously difficult. Examples of the application of high-performance liquid chromatography (HPLC) with the improved OR detector include (i) the analytical scale separation of fructose and sucrose and (ii) the semi-preparative separation of enantiomers of warfarin and Trögers base. A signal-to-noise improvement of up to 150% is found when comparing signals with and without correction for transmittance changes. The improved OR detector has been used in series with a UV detector and the system shown to be suitable for on-line measurement of peak purity in separations using a chiral column under overload conditions.