Suppr超能文献

圆锥形、锥形种植体基台的拆除扭矩:阳极氧化和表面积减小的影响。

Removal torques of conical, tapered implant abutments: the effects of anodization and reduction of surface area.

作者信息

Squier Rachel S, Psoter Walter J, Taylor Thomas D

机构信息

Department of Prosthodontics and Operative Dentistry, University of Connecticut, School of Dental Medicine, Farmington 06030-1615, USA.

出版信息

Int J Oral Maxillofac Implants. 2002 Jan-Feb;17(1):24-7.

Abstract

PURPOSE

To examine the effects of anodization (surface coating) and reduction of internal Morse taper surface area on the reversal torque values of Straumann ITI dental implants and abutments.

MATERIALS AND METHODS

Eighty ITI solid screw implants were mated with corresponding 5.5-mm solid abutments. The assemblies were divided into 4 test groups of 20 specimens. All abutments were torque tightened into the implant to 35 Ncm. Half of the abutments were anodized and half were in their as-machined state. Each of these 2 groups included half of the implants with the standard internal Morse taper configuration and half with the synOcta (Straumann USA, Waltham, MA) internal positioning interface (indexed). Torque removal testing was then performed on the assemblies. The 4 groups were compared statistically to examine the effect of the 2 variables (anodization and reduction in surface area). Scheffe's test for multiple comparisons was used to compare groups at an adjusted significance level of < or = .05.

RESULTS

Torque removal of all specimens revealed that the indexed implant with the non-anodized abutment demonstrated superior removal torque.

DISCUSSION

The indexed and standard implants with anodized abutments, and the standard implant with the non-anodized abutment had lower reversal torque values.

CONCLUSION

The addition of the indexed internal surface to the ITI implant did not have deleterious effect on the resistance to loosening of standard solid abutments.

摘要

目的

研究阳极氧化(表面涂层)和减小内部莫氏锥度表面积对士卓曼ITI牙科种植体及基台的反转扭矩值的影响。

材料与方法

80颗ITI实心螺钉种植体与相应的5.5毫米实心基台配对。将组件分为4个测试组,每组20个样本。所有基台均以35 Ncm的扭矩拧紧至种植体内。一半基台进行阳极氧化处理,另一半保持加工后的状态。这两组中的每组都包括一半具有标准内部莫氏锥度配置的种植体和一半具有synOcta(美国士卓曼公司,马萨诸塞州沃尔瑟姆)内部定位界面(带索引)的种植体。然后对组件进行扭矩移除测试。对这4组进行统计学比较,以研究两个变量(阳极氧化和表面积减小)的影响。使用谢费多重比较检验在调整后的显著性水平≤0.05下比较各组。

结果

所有样本的扭矩移除显示,带有未阳极氧化基台的带索引种植体表现出更高的移除扭矩。

讨论

带有阳极氧化基台的带索引种植体和标准种植体,以及带有未阳极氧化基台的标准种植体,其反转扭矩值较低。

结论

在ITI种植体上增加带索引的内表面,对标准实心基台的抗松动性没有有害影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验