Optimal deployment of third-generation stents: an intravascular ultrasound assessment.

作者信息

Maddux James T, Carroll John D, Groves Bertron M, Messenger John C, Tseng Albert, Falcone Erin, Burchenal J E B

机构信息

University of Colorado Health Sciences Center, Denver, Colorado 80262, USA.

出版信息

Catheter Cardiovasc Interv. 2002 Oct;57(2):142-7. doi: 10.1002/ccd.10302.

Abstract

Third-generation intracoronary stents allow deployment at higher pressures, possibly obviating the need for high-pressure postdilations and also possibly reducing restenosis. This study evaluated the ability of the Tristar Coronary Stent System to produce optimal stent deployment as measured by intravascular ultrasound (IVUS) and quantitative coronary angiography in 46 patients. Optimal stent deployment was defined as minimal luminal area > 80% of the average of the proximal and distal reference luminal areas. After initial deployment, 74.5% of stents met criteria for optimal stent deployment by IVUS, with an average stent expansion ratio of 89.6%. Ten stents (18.2%) were postdilated. Four patients (8.7%) had a major adverse cardiac event, one patient died, one patient had a myocardial infarction, and two patients had target vessel revascularization at 6 months. The Tristar stent system produces optimal deployment without the need for routine postdilation and results in optimal clinical outcomes.