Darwin Ling H, Cullen Allison P, Crowe Susanne R, Modarress Kevin J, Willis Dean E, Payne William J
Digene Corporation, Gaithersburg, Maryland 20878, USA.
Sex Transm Dis. 2002 Oct;29(10):576-80. doi: 10.1097/00007435-200210000-00003.
A previous study demonstrated that Digene's Hybrid Capture 2 (HC2) DNA tests for detection of and (CT/GC) could be performed using cervical swab specimens collected in GenProbe transport media with significantly greater sensitivity for the detection of than with the GenProbe PACE 2 system.
The goal was to assess the performance of HC2 tests in comparison with GenProbe PACE 2 tests for the detection of CT/GC in male urethral swab specimens.
A total of 1,202 male urethral swab specimens were collected in GenProbe PACE transport medium. All specimens were first tested with the PACE 2 system, followed by masked HC2 CT/GC testing. The GenProbe AMPLIFIED CT Assay (AMP CT) and PCR/SHARP Signal System (SHARP) were used for adjudication of discrepant results.
The prevalence rates for this population were 8.4% for and 14.6% for, based on the adjudicated results. The relative sensitivity and specificity for the detection of were 97.0% and 99.8% for HC2 and 69.3% and 98.3% for PACE 2, respectively. The relative sensitivities for the detection of were 98.9% for HC2 and 99.4% for PACE 2, with the same specificity of 99.9% for both tests. Agreement between the two testing methods was 95.4% for and 99.6% for.
The HC2 test is compatible with the GenProbe collection medium, with significantly greater sensitivity than the GenProbe PACE 2 test for detecting and similar sensitivities for detecting.
先前的一项研究表明,Digene公司的杂交捕获2(HC2)DNA检测方法用于检测沙眼衣原体(CT)和淋病奈瑟菌(GC)时,可使用收集于GenProbe转运培养基中的宫颈拭子标本进行检测,其检测CT的灵敏度显著高于GenProbe PACE 2系统。
目的是评估HC2检测方法与GenProbe PACE 2检测方法在检测男性尿道拭子标本中CT/GC的性能。
总共收集了1202份男性尿道拭子标本,并置于GenProbe PACE转运培养基中。所有标本首先用PACE 2系统进行检测,随后进行盲法HC2 CT/GC检测。使用GenProbe扩增CT检测法(AMPLIFIED CT Assay,AMP CT)和PCR/SHARP信号系统(SHARP)来判定不一致的结果。
根据判定结果,该人群中CT的患病率为8.4%,GC的患病率为14.6%。HC2检测CT的相对灵敏度和特异性分别为97.0%和99.8%,PACE 2检测CT的相对灵敏度和特异性分别为69.3%和98.3%。HC2检测GC的相对灵敏度为98.9%,PACE 2检测GC的相对灵敏度为99.4%,两种检测方法检测GC的特异性均为99.9%。两种检测方法对CT检测结果的一致性为95.4%,对GC检测结果的一致性为99.6%。
HC2检测方法与GenProbe收集培养基兼容,检测CT时的灵敏度显著高于GenProbe PACE 2检测方法,但检测GC时的灵敏度相似。