[Visual user guidance for registration based on A-mode ultrasound].

作者信息

Heger S, Portheine F, Ohnsorge J, Radermacher K

机构信息

Institut für Biomedizinische Technologien, RWTH Aachen, Deutschland.

出版信息

Biomed Tech (Berl). 2002;47 Suppl 1 Pt 1:61-4. doi: 10.1515/bmte.2002.47.s1a.61.

Abstract

In this study we investigate the use of an A-Mode ultrasound probe tracked by a mechanical localizer system (MicroScribe 3D) for non-invasive transcutaneous palpation and registration of bone surface points. A CT of a femur-model has been scanned and subsequently processed by segmentation and 3D-reconstruction. During the ultrasound registration, a computer based assistance tool helped guiding the alignment of the ultrasound probe. Three different modes of registration have been compared. Process times for registration have been recorded and compared. The results for using an A-Mode ultrasound system demonstrate a translation RMS accuracy of 0.58 mm. Mean ultrasound registration time has been measured to 108 sec. for palpation of 10 bone surface points of distal part of femur.

摘要

在本研究中,我们研究了一种由机械定位系统(MicroScribe 3D)跟踪的A模式超声探头用于无创经皮触诊和骨表面点配准的情况。对一个股骨模型进行了CT扫描,随后通过分割和三维重建进行处理。在超声配准过程中,一个基于计算机的辅助工具帮助引导超声探头的对准。比较了三种不同的配准模式。记录并比较了配准的处理时间。使用A模式超声系统的结果表明平移均方根误差精度为0.58毫米。测量得到平均超声配准时间为108秒,用于触诊股骨远端的10个骨表面点。