BACKGROUND

Acylcarnitine (AC) profiling in dried blood spots by means of electrospray ionisation tandem mass spectrometry (ESI-MS/MS) has proven to be a useful method in neonatal screening, able to detect inborn errors of fatty acid oxidation, amino acid, organic acid and carnitine metabolism. Furthermore, this method is becoming increasingly applied in selective screening and in prenatal and postmortal diagnostics of inborn metabolic disorders, where urine is commonly used as specimen of interest. We therefore developed and validated a butylation method of acylcarnitine profiling in urine by ESI-MS/MS without previous chromatographic separation.

METHODS

Random urine specimens were used for investigation of the analytical imprecision of the method. Recovery, precision and linearity were determined using methanolic standard solutions of free carnitine, octanoylcarnitine and palmitoylcarnitine at various concentrations.

RESULTS

The mean coefficients of variation of within-run and run-to-run analysis of these analytes were found between 10% and 20% and demonstrated that the method fulfills the analytical requirements within the relevant ranges of concentration. Creatinine-related and age-related reference values of free carnitine and the ACs (C2-C18) were established. The definite discrimination was possible between patients with fatty acid oxidation disorders, organic acidurias, and healthy controls. The AC profiles from patients with various specific disorders were diagnostically helpful during acute deterioration and even during conditions of well-compensated metabolic state.

CONCLUSION

The method used in this study is suitable both for selective screening and for confirmation of diagnosis with the advantage of high-throughput quantitative measurement.

摘要

背景

通过电喷雾电离串联质谱法(ESI-MS/MS)对干血斑中的酰基肉碱(AC)进行分析,已被证明是新生儿筛查中的一种有用方法,能够检测脂肪酸氧化、氨基酸、有机酸和肉碱代谢的先天性缺陷。此外,这种方法越来越多地应用于先天性代谢紊乱的选择性筛查以及产前和死后诊断,在这些诊断中,尿液通常作为感兴趣的样本。因此,我们开发并验证了一种无需先前色谱分离的ESI-MS/MS法对尿液中的酰基肉碱进行丁基化分析的方法。

方法

使用随机尿液样本研究该方法的分析不精密度。使用不同浓度的游离肉碱、辛酰肉碱和棕榈酰肉碱的甲醇标准溶液测定回收率、精密度和线性。

结果

这些分析物批内和批间分析的平均变异系数在10%至20%之间,表明该方法在相关浓度范围内满足分析要求。建立了游离肉碱和ACs(C2-C18)与肌酐相关和年龄相关的参考值。脂肪酸氧化障碍患者、有机酸尿症患者和健康对照之间能够进行明确区分。在急性病情恶化甚至代谢状态良好代偿的情况下,各种特定疾病患者的AC谱对诊断都有帮助。

结论

本研究中使用的方法适用于选择性筛查和诊断确认,具有高通量定量测量的优势。