电极表面积和刺激阈值在永久性心脏起搏中的意义。

The significance of electrode surface area and stimulating thresholds in permanent cardiac pacing.

作者信息

Smyth N P, Tarjan P P, Chernoff E, Baker N

出版信息

J Thorac Cardiovasc Surg. 1976 Apr;71(4):559-65.

PMID:1263538
Abstract

This paper reports a long-term clinical study of evolving chronic thresholds of cardiac stimulation in patients with implanted pacemakers and two types of permanent prevenous unipolar pacing leads. A comprehensive collection of threshold measurements is presented and analyzed. These data were accumulated over a period of more than 4 years by measuring stimulation thresholds at each of five stimulus pulse durations ranging from 0.5 to 2.5 msec. Measurements were made at initial implantation and at first (21 month) and second (42 month) reimplantations.

摘要

本文报告了一项针对植入起搏器及两种类型的永久性单极静脉起搏导线患者心脏刺激慢性阈值演变的长期临床研究。文中展示并分析了阈值测量的全面数据收集情况。这些数据是通过在0.5至2.5毫秒的五个刺激脉冲持续时间下测量刺激阈值,历经4年多时间积累而成。测量在初次植入时以及首次(21个月)和第二次(42个月)重新植入时进行。