中耳植入物——Symphonix 活力听觉桥:对法国首批125例植入患者的回顾性研究

A middle ear implant, the Symphonix Vibrant Soundbridge: retrospective study of the first 125 patients implanted in France.

作者信息

Sterkers Olivier, Boucarra Didier, Labassi Samia, Bebear Jean-Pierre, Dubreuil Christian, Frachet Bruno, Fraysse Bernard, Lavieille Jean-Pierre, Magnan Jacques, Martin Christian, Truy Eric, Uziel Alain, Vaneecloo François Michel

机构信息

Hôpital Beaujon AP-HP, INSERM EMI-U0112, Faculté Xavier Bichat, Université Paris, Clichy, France.

出版信息

Otol Neurotol. 2003 May;24(3):427-36. doi: 10.1097/00129492-200305000-00013.

Abstract

OBJECTIVE

The Vibrant Soundbridge is an active semi-implantable middle ear implant for the rehabilitation of patients with a sensorineural hearing loss who are not able to derive adequate benefit from conventional hearing aids. A retrospective study was performed to assess the overall level of satisfaction of implanted patients and to investigate the potential determinants of postoperative success.

STUDY DESIGN

A retrospective survey of audiological data from repeated measures and subjective data from self-assessment scales administered postoperatively was conducted to determine the degree of benefit and satisfaction for Vibrant Soundbridge implantees.

SETTING

Twenty-one tertiary referral and teaching hospitals.

SUBJECTS

The first 125 VSB implantees implanted in France between August 1997 and May 2001 were included in the study.

RESULTS

No clinically significant change was observed for residual hearing postoperatively. Most patients (83%) reported they were either satisfied or very satisfied with the Vibrant Soundbridge. Analysis of correlation revealed a moderate correlation (Pearson coefficient r = 0.59) between the degree of benefit reported via the patient survey and the degree of benefit reported via the Glasgow Benefit Inventory. A moderate correlation (Pearson coefficient r = 0.66) was observed between speech comprehension scores in quiet for the preoperative unaided condition and the postoperative aided Vibrant Soundbridge condition. No correlation was observed between subjective reports of satisfaction postoperatively and performance on preoperative objective tests or patient characteristics.

CONCLUSION

The results indicate a high level of satisfaction with the VSB as a treatment of sensorineural hearing impairment in patients with a wide range of characteristics. Preoperative scores for unaided speech comprehension tests in quiet may be a potential indicator of success on aided Vibrant Soundbridge speech comprehension tests postoperatively but do not reflect patient satisfaction with the device reported on self-assessment scales.

摘要

目的

Vibrant Soundbridge是一种主动式半植入式中耳植入装置,用于帮助感音神经性听力损失患者康复,这些患者无法从传统助听器中获得足够益处。本研究进行了一项回顾性研究,以评估植入患者的总体满意度,并探究术后成功的潜在决定因素。

研究设计

对重复测量的听力学数据以及术后自我评估量表的主观数据进行回顾性调查,以确定Vibrant Soundbridge植入者的受益程度和满意度。

研究地点

21家三级转诊和教学医院。

研究对象

纳入了1997年8月至2001年5月在法国首次植入125例Vibrant Soundbridge的患者。

结果

术后残余听力未观察到具有临床意义的变化。大多数患者(83%)报告对Vibrant Soundbridge感到满意或非常满意。相关性分析显示,通过患者调查报告的受益程度与通过格拉斯哥受益量表报告的受益程度之间存在中度相关性(皮尔逊系数r = 0.59)。术前未佩戴助听器时安静环境下的言语理解得分与术后佩戴Vibrant Soundbridge时的言语理解得分之间观察到中度相关性(皮尔逊系数r = 0.66)。术后主观满意度报告与术前客观测试表现或患者特征之间未观察到相关性。

结论

结果表明,对于具有广泛特征的患者,作为感音神经性听力障碍的一种治疗方法,Vibrant Soundbridge的满意度较高。术前安静环境下未佩戴助听器时的言语理解测试得分可能是术后佩戴Vibrant Soundbridge时言语理解测试成功的一个潜在指标,但不能反映患者在自我评估量表上报告的对该装置的满意度。