Catheter Cardiovasc Interv. 2003 Aug;59(4):496-503. doi: 10.1002/ccd.10583.
The efficacy and safety of coronary stent implantation depend on the mechanical features of these devices when deployed in atheromatous lesions of various morphologies. We evaluated the trackability, flexibility, and conformability of 17 coronary stents using specific mechanical bench tests. The quantifications used a dynamometer for assessment of trackability (maximal strength) and flexibility (stiffness) and a 3D optical gauging machine for assessment of conformability (distance between stent and arterial wall in a curvature). The maximal strength (measuring the trackability) ranged respectively from 0.24 +/- 0.06 and 0.38 +/- 0.03 N (Seaquest) to 1.31 +/- 0.42 and 1.34 +/- 0.35 N (Carbostent), concerning respectively curvatures of 90 degrees (P < 0.0001) and 135 degrees (P < 0.0001). The stiffness (measuring the flexibility) ranged from 0.53 +/- 0.16 (Seaquest) to 1.28 +/- 0.10 N/mm (NIR Royal; P < 0.0001). The mean distance between stent and external curvature (external conformability) ranged from 0.15 +/- 0.06 mm (S7) to 0.57 +/- 0.4 mm (NIR Royal; P < 0.0001). The mean distance between stent and internal curve (internal conformability) ranged from 0.26 +/- 0.13 (S7) to 0.44 +/- 0.12 mm (S670; P < 0.0001). These results may influence the choice of a particular stent adapted to a specific coronary anatomy.
冠状动脉支架植入的有效性和安全性取决于这些装置在各种形态的动脉粥样硬化病变中展开时的机械特性。我们使用特定的机械台架试验评估了17种冠状动脉支架的可跟踪性、柔韧性和顺应性。量化过程中,使用测力计评估可跟踪性(最大强度)和柔韧性(刚度),使用三维光学测量仪评估顺应性(支架在弯曲处与动脉壁之间的距离)。最大强度(测量可跟踪性)在90度曲率(P<0.0001)和135度曲率(P<0.0001)时,分别从0.24±0.06和0.38±0.03N(Seaquest)到1.31±0.42和1.34±0.35N(Carbostent)。刚度(测量柔韧性)范围从0.53±0.16(Seaquest)到1.28±0.10N/mm(NIR Royal;P<0.0001)。支架与外部曲率之间的平均距离(外部顺应性)范围从0.15±0.06mm(S7)到0.57±0.4mm(NIR Royal;P<0.0001)。支架与内部曲线之间的平均距离(内部顺应性)范围从0.26±0.13(S7)到0.44±0.12mm(S670;P<0.0001)。这些结果可能会影响针对特定冠状动脉解剖结构选择特定的支架。