Aspradakis Maria M, Morrison Rachel H, Richmond Neil D, Steele Alasdair
Regional Medical Physics Department, Newcastle General Hospital, Westgate Road, Newcastle upon Tyne, NE4 6BE, UK.
Phys Med Biol. 2003 Sep 7;48(17):2873-93. doi: 10.1088/0031-9155/48/17/309.
This work describes an experimental verification of the two-photon dose calculation engines available on the Helax-TMS (version 6.1) commercial radiotherapy treatment planning system. The performance of the pencil beam convolution and the collapsed cone superposition algorithms was examined for 4, 6, 15 MV beams, under a range of clinically relevant irradiation geometries. Comparisons against measurements were carried out in terms of absolute dose, thus assessment of the accuracy of monitor unit (MU) calculations was also carried out. Results show that both algorithms agree with measurement to acceptable tolerance levels in most cases in homogeneous water-equivalent media irradiated under full scatter conditions. The collapsed cone algorithm slightly overestimates the penumbra width and this is mainly due to discretization effects of the fluence matrix. The accuracy of this algorithm strongly depends on the resolution of the patient density matrix. It is recommended that the density matrix voxel size used for dose calculations is less than 5 x 5 x 5 mm3. The dose in media irradiated under missing tissue geometry, or in the presence of low or high-density heterogeneities, is modelled best with the collapsed cone algorithm. This is of particular clinical interest in treatment planning of the breast and of the thorax. For these treatment sites, a retrospective study of treatment plans indicated in certain cases significant overestimation of the dose to the planning target volume when using the pencil beam convolution model.
这项工作描述了对Helax-TMS(版本6.1)商业放射治疗治疗计划系统上可用的双光子剂量计算引擎的实验验证。在一系列临床相关的照射几何条件下,对4、6、15兆伏射线束的笔形束卷积和坍缩锥叠加算法的性能进行了研究。针对绝对剂量与测量值进行了比较,因此还对监测单位(MU)计算的准确性进行了评估。结果表明,在全散射条件下照射的均匀水等效介质中,在大多数情况下,两种算法与测量值的偏差都在可接受的公差范围内。坍缩锥算法略微高估了半值层宽度,这主要是由于注量矩阵的离散化效应。该算法的准确性强烈依赖于患者密度矩阵的分辨率。建议用于剂量计算的密度矩阵体素大小小于5×5×5立方毫米。在缺失组织几何条件下照射的介质中,或在存在低密度或高密度不均匀性的情况下,使用坍缩锥算法对剂量的建模效果最佳。这在乳房和胸部的治疗计划中具有特别的临床意义。对于这些治疗部位,一项治疗计划的回顾性研究表明,在某些情况下,使用笔形束卷积模型时,对计划靶体积的剂量有显著高估。