Initial experience with a new 8 French-compatible directional atherectomy catheter: immediate and mid-term results.

作者信息

Orlic Dejan, Reimers Bernhard, Stankovic Goran, Corvaja Nicola, Chieffo Alaide, Airoldi Flavio, Spanos Vassilis, Favero Luca, Di Mario Carlo, Colombo Antonio

机构信息

Catheterization Laboratory, Columbus Hospital, Milan, Italy.

出版信息

Catheter Cardiovasc Interv. 2003 Oct;60(2):159-66. doi: 10.1002/ccd.10633.

Abstract

The purpose of this study was to evaluate the safety and efficacy of the new Fox Hollow atherectomy device (FHT) designed for more efficient and easier plaque removal. The FHT has short rigid section and low-profile cutter mounted on a monorail catheter. The FHT catheter was utilized in 77 patients with 98 lesions. Mean reference vessel diameter was 2.75 +/- 0.51 mm. Successful atherectomy with tissue retrieval was performed in 94 lesions (96%). Following atherectomy, mean diameter stenosis was reduced from 71.1% to 31.9% and further to 10.4% following adjunctive treatment. Angiographic complications were one coronary perforation and one adventitial staining, both successfully treated with prolong balloon inflation and stent implantation. Nine patients (11.7%) had in-hospital non-Q-wave myocardial infarction (MI). One patient died (1.3%) for noncardiac reasons and one had MI (1.3%) at 6-month follow-up. Target lesion revascularization was required in 13 (13.8%) lesions and target vessel revascularization in 15 (20.3%) patients. There was target vessel failure in 17 (23.0%) patients. Plaque debulking with the FHT catheter can be performed safely and effectively in relatively small vessels and complex lesions located in mid-distal artery segments with 6-month clinical outcome similar to prior atherectomy devices.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验