Stamatelatos I E, Kasviki K, Green S, Gainey M, Kalef-Ezra J, Beddoe A
Institute of Nuclear Technology and Radiation Protection, NCSR Demokritos, Aghia Paraskevi, Greece.
Anal Bioanal Chem. 2004 May;379(2):192-7. doi: 10.1007/s00216-003-2460-6. Epub 2004 Feb 5.
The design, calibration, dosimetry and performance evaluation of a prompt-gamma neutron activation analysis facility for in vivo body composition studies in small animals (i.e. rats or rabbits) is discussed. The system design was guided by Monte Carlo transport calculations using MCNP-4C code. A system was built and performance evaluation was made using a 185-GBq Pu-Be neutron source. Prompt-gamma rays produced by neutron capture reactions were detected by a combination of a NaI(Tl) scintillation and a HPGe semiconductor detectors. Nitrogen and chlorine were quantified by analysis of the 10.83-MeV and 6.11-MeV peaks, respectively. Appropriate corrections for the animal body size were determined. The facility described allows the in vivo determination of protein and extracellular space in sets of experimental animals.
讨论了用于小动物(即大鼠或兔子)体内成分研究的瞬发伽马中子活化分析设备的设计、校准、剂量测定和性能评估。系统设计以使用MCNP-4C代码的蒙特卡罗输运计算为指导。构建了一个系统,并使用185 GBq的钚铍中子源进行了性能评估。通过碘化钠(铊)闪烁探测器和高纯锗半导体探测器组合检测中子俘获反应产生的瞬发伽马射线。分别通过分析10.83 MeV和6.11 MeV峰对氮和氯进行定量。确定了对动物体型的适当校正。所述设备允许在实验动物组中体内测定蛋白质和细胞外空间。