Ankrum Matthew T, Hartwell Gary R, Truitt John E
Department of Endodontics, Virginia Commonwealth University School of Dentistry, Richmond, USA.
J Endod. 2004 Apr;30(4):234-7. doi: 10.1097/00004770-200404000-00013.
It was the aim of this study to investigate the incidence of file breakage and distortion when the ProTaper, K3 Endo, and ProFile systems were used to instrument canals in the severely curved roots of extracted molars. Forty-five roots of extracted mandibular and maxillary molars with curvatures between 40 and 75 degrees were chosen for this study. The canals in group 1 were instrumented with the ProFile system. Roots in group 2 were instrumented with the ProTaper system, and those in group 3 were instrumented with the K3 Endo system. The three systems were used according to the manufacturers' instructions. The proportion of files distorted was 15.3% for the ProFile group, 2.4% for the ProTaper group, and 8.3% for the K3 Endo group. There was a statistically significant difference between the ProFile and ProTaper groups (p=0.0079). The percentage of broken files was 1.7% for the ProFile group, 6.0% for the ProTaper group, and 2.1% for the K3 Endo group. No statistically significant differences were found between these three groups (p=0.4243). The results of this study showed that these three rotary tapered systems were not significantly different with regard to breakage. There were significantly more distorted files in the ProFile group when compared with the ProTaper group. With regard to distortion, there was no significant difference between the ProTaper and K3 Endo and the ProFile and K3 Endo groups.
本研究旨在调查使用ProTaper、K3 Endo和ProFile系统对拔除磨牙的严重弯曲根管进行预备时锉针折断和变形的发生率。本研究选取了45颗弯曲度在40至75度之间的拔除下颌和上颌磨牙的牙根。第1组的根管使用ProFile系统进行预备。第2组的牙根使用ProTaper系统进行预备,第3组的牙根使用K3 Endo系统进行预备。这三种系统均按照制造商的说明使用。ProFile组锉针变形的比例为15.3%,ProTaper组为2.4%,K3 Endo组为8.3%。ProFile组和ProTaper组之间存在统计学显著差异(p = 0.0079)。ProFile组锉针折断的百分比为1.7%,ProTaper组为6.0%,K3 Endo组为2.1%。这三组之间未发现统计学显著差异(p = 0.4243)。本研究结果表明,这三种旋转锥形系统在折断方面无显著差异。与ProTaper组相比,ProFile组中变形的锉针明显更多。在变形方面,ProTaper组与K3 Endo组以及ProFile组与K3 Endo组之间无显著差异。