Rassiah P, Ng K H, DeWerd L A, Kunugi K
Clinical Oncology Unit, University of Malaya Medical Centre, Kuala Lumpur, Malaysia.
Australas Phys Eng Sci Med. 2004 Mar;27(1):25-9. doi: 10.1007/BF03178885.
A thermoluminescent dosimetry (TLD) postal dose inter-comparison was carried out amongst radiotherapy centres in Malaysia. The aim of this TLD inter-comparison was to compare the uniformity involved in the measurement of absorbed dose among the participating centres. A set of 5 TLD chips placed within acrylic trays were mailed to all participating centres for irradiation to an absorbed dose to water of 2 Gy. Measurements were made for 6 MV and 60Co photon beams. Results show an agreement of +/- 5% for all but three radiotherapy centres. The ratios of the TLD readings to that of the reference centre are comparable with other national/regional dose inter-comparisons. The importance of a proper ongoing quality assurance program is essential in maintaining the consistency and uniformity of doses delivered.