治疗场与脉冲多普勒场中超声暴露的比较。

A comparison of ultrasound exposure in therapy and pulsed Doppler fields.

作者信息

Starritt H C, Duck F A

机构信息

Department of Medical Physics, Royal United Hospital, Bath, Avon, UK.

出版信息

Br J Radiol. 1992 Jul;65(775):557-63. doi: 10.1259/0007-1285-65-775-557.

Abstract

A detailed comparison of the ultrasound exposure in water from a therapeutic beam and a pulsed Doppler beam was carried out. A significant overlap in acoustic power was found between therapy intensity levels used clinically and the upper end of the diagnostic range, between approximately 100 mW and 200 mW. In addition, pulse pressure amplitudes in the range 0.5-1.0 MPa were measured close to the transducer on both units. It is common to use physiotherapy equipment at pulse average intensities of 0.5 W/cm2 or less, and at these levels exposures of similar magnitude may be obtained with beams currently defined as therapeutic and those available from pulsed Doppler equipment.

摘要

对治疗束和脉冲多普勒束在水中的超声暴露进行了详细比较。发现在临床使用的治疗强度水平与诊断范围上限之间,即大约100毫瓦至200毫瓦之间,声功率存在显著重叠。此外,在两个设备上靠近换能器处测量到的脉冲压力幅度在0.5 - 1.0兆帕范围内。通常使用脉冲平均强度为0.5瓦/平方厘米或更低的物理治疗设备,在这些水平下,目前定义为治疗性的束流和脉冲多普勒设备产生的束流可能会导致类似程度的暴露。