Kaczmarek K A, Webster J G, Radwin R G
Department of Electrical and Computer Engineering, University of Wisconsin, Madison 53706.
IEEE Trans Biomed Eng. 1992 Jul;39(7):701-15. doi: 10.1109/10.142645.
A new method to measure the dynamic range of electrotactile (electrocutaneous) stimulation uses both steepest ascent (gradient) and one-variable-at-a-time methods to determine the waveform variables that maximize the subjective magnitude (intensity) of the electrotactile percept at the maximal current without discomfort for balanced-biphasic pulse bursts presented at a 15-Hz rate. The magnitude at the maximal current without discomfort is maximized by the following waveform (range tested in parentheses): number of pulses/burst = 6 (1-20), pulse repetition rate within a burst = 350 Hz (200-1500), and phase width = 150 microseconds (40-350). The interphase interval (separation between positive and negative phases in a biphasic pulse) does not affect dynamic range from 0-500 microseconds. The number of pulses/burst has a large effect on the perceived dynamic range when this is measured using a subjective-magnitude-based algorithm, whereas it has little effect on the traditional dynamic range measure, i.e., (maximal current without discomfort)/(sensation threshold current). The perceived stimulus magnitude at the maximal current without discomfort is approximately twice as strong with 6 pulses/burst as it is with 1 pulse/burst (a frequently-used waveform).
一种测量电触觉(电皮肤)刺激动态范围的新方法,使用最速上升法(梯度法)和一次改变一个变量的方法,来确定在以15赫兹频率呈现的平衡双相脉冲串的最大电流下,能使电触觉感知的主观强度最大化且不会引起不适的波形变量。通过以下波形(括号内为测试范围)可使在最大电流下无不适时的强度最大化:每串脉冲数 = 6(1 - 20),一串内的脉冲重复率 = 350赫兹(200 - 1500),以及相位宽度 = 150微秒(40 - 350)。双相脉冲中相间间隔(正负相位之间的间隔)在0 - 500微秒范围内不影响动态范围。当使用基于主观强度的算法测量时,每串脉冲数对感知到的动态范围有很大影响,而对传统的动态范围测量方法,即(最大无不适电流)/(感觉阈值电流)影响很小。在最大电流下无不适时,每串6个脉冲的感知刺激强度大约是每串1个脉冲(一种常用波形)时的两倍。