Time-domain cross-correlation baroreflex sensitivity: performance on the EUROBAVAR data set.

作者信息

Westerhof Berend E, Gisolf Janneke, Stok Wim J, Wesseling Karel H, Karemaker John M

机构信息

TNO-TPD-Biomedical Instrumentation, Academic Medical Centre, University of Amsterdam, The Netherlands.

出版信息

J Hypertens. 2004 Jul;22(7):1371-80. doi: 10.1097/01.hjh.0000125439.28861.ed.

Abstract

OBJECTIVE

To test a new method (cross-correlation baroreflex sensitivity, xBRS) for the computation of time-domain baroreflex sensitivity on spontaneous blood pressure and heart interval variability using the EUROBAVAR data set.

METHODS

We applied xBRS to the 42 records in the EUROBAVAR data set, obtained from 21 patients in the lying and standing positions. One patient had a recent heart transplant and one was diabetic with evident cardiac autonomic neuropathy. xBRS computes the correlation between beat-to-beat systolic blood pressure and R-R interval, resampled at 1 Hz, in a sliding 10 s window, with delays of 0-5 s for interval. The delay with the greatest positive correlation is selected and, when significant at P = 0.01, slope and delay are recorded as one xBRS value. Each 1 s of the recording is the start of a new computation. Non-parametric tests are used.

RESULTS

With patients in the lying position, xBRS yielded a value of 12.4 ms/mmHg compared with the EUROBAVAR sequential 16.2 ms/mmHg, and for the standing positions the respective values were 6.2 and 6.7 ms/mmHg, giving lying to standing ratios of 1.96 and 2.10, respectively. xBRS yielded results for all files, with 20 values per minute on average at a lower within-patient variance. Best delays were 0, 1 and 2 s, and the delay increased by 102 ms when the patient was in the standing position. The xBRS method was successful in the patients with diabetes and the heart transplant.

CONCLUSION

The xBRS method should be considered for experimental and clinical use, because it yielded values that correlated strongly with and were close to the EUROBAVAR averages, yielded more values per minute, had lower within-patient variance and measured baroreflex delay.

摘要

目的

使用EUROBAVAR数据集,测试一种用于计算基于自发性血压和心率间期变异性的时域压力反射敏感性的新方法(互相关压力反射敏感性,xBRS)。

方法

我们将xBRS应用于EUROBAVAR数据集中的42条记录,这些记录来自21名处于卧位和立位的患者。其中一名患者近期接受了心脏移植,一名患者患有糖尿病且伴有明显的心脏自主神经病变。xBRS计算逐搏收缩压与R-R间期之间的相关性,在1 Hz下重采样,在10 s的滑动窗口中进行,间期延迟为0至5 s。选择具有最大正相关性的延迟,当P = 0.01时具有显著性时,斜率和延迟记录为一个xBRS值。记录的每1 s都是新计算的开始。使用非参数检验。

结果

患者处于卧位时,xBRS值为12.4 ms/mmHg,而EUROBAVAR序列法为16.2 ms/mmHg;立位时,相应值分别为6.2和6.7 ms/mmHg,卧位与立位比值分别为1.96和2.10。xBRS对所有文件都得出了结果,平均每分钟有20个值,患者内方差较低。最佳延迟为0、1和2 s,患者站立时延迟增加102 ms。xBRS方法在糖尿病患者和心脏移植患者中均成功。

结论

xBRS方法应考虑用于实验和临床,因为它得出的值与EUROBAVAR平均值高度相关且接近,每分钟得出的值更多,患者内方差更低,并测量了压力反射延迟。