Suppr超能文献

用于四维成像的256层CT扫描仪的物理性能评估。

Physical performance evaluation of a 256-slice CT-scanner for four-dimensional imaging.

作者信息

Mori Shinichiro, Endo Masahiro, Tsunoo Takanori, Kandatsu Susumu, Tanada Shuji, Aradate Hiroshi, Saito Yasuo, Miyazaki Hiroaki, Satoh Kazumasa, Matsushita Satoshi, Kusakabe Masahiro

机构信息

National Institute of Radiological Sciences, Chiba 263-8555, Japan.

出版信息

Med Phys. 2004 Jun;31(6):1348-56. doi: 10.1118/1.1747758.

Abstract

We have developed a prototype 256-slice CT-scanner for four-dimensional (4D) imaging that employs continuous rotations of a cone-beam. Since a cone-beam scan along a circular orbit does not collect a complete set of data to make an exact reconstruction of a volume [three-dimensional (3D) image], it might cause disadvantages or artifacts. To examine effects of the cone-beam data collection on image quality, we have evaluated physical performance of the prototype 256-slice CT-scanner with 0.5 mm slices and compared it to that of a 16-slice CT-scanner with 0.75 mm slices. As a result, we found that image noise, uniformity, and high contrast detectability were independent of z coordinate. A Feldkamp artifact was observed in distortion measurements. Full width at half maximum (FWHM) of slice sensitivity profiles (SSP) increased with z coordinate though it seemed to be caused by other reasons than incompleteness of data. With regard to low contrast detectability, smaller objects were detected more clearly at the midplane (z = 0 mm) than at z = 40 mm, though circular-band like artifacts affected detection. The comparison between the 16-slice and the 256-slice scanners showed better performance for the 16-slice scanner regarding the SSP, low contrast detectability, and distortion. The inferiorities of the 256-slice scanner in other than distortion measurement (Feldkamp artifact) seemed to be partly caused by the prototype nature of the scanner and should be improved in the future scanner. The image noise, uniformity, and high contrast detectability were almost identical for both CTs. The 256-slice scanner was superior to the 16-slice scanner regarding the PSF, though it was caused by the smaller transverse beam width of the 256-slice scanner. In order to compare both scanners comprehensively in terms of exposure dose, noise, slice thickness, and transverse spatial resolution, K=Dsigma2ha3 was calculated, where D was exposure dose (CT dose index), sigma was magnitude of noise, h was slice thickness (FWHM of SSP), and a was transverse spatial resolution (FWHM of PSF). The results showed that the K value was 25% larger for the 16-slice scanner, and that the 256-slice scanner was 1.25 times more effective than the 16-slice scanner at the midplane. The superiority in K value for the 256-slice scanner might be partly caused by decrease of wasted exposure with a wide-angle cone-beam scan. In spite of the several problems of the 256-slice scanner, it took a volume data approximately 1.0 mm (transverse) x 1.3 mm (longitudinal) resolution for a wide field of view (approximately 100 mm long) along the zeta axis in a 1 s scan if resolution was defined by the FWHM of the PSF or the SSP, which should be very useful to take dynamic 3D (4D) images of moving organs.

摘要

我们开发了一种用于四维(4D)成像的256层CT扫描仪原型,该扫描仪采用锥形束的连续旋转。由于沿圆形轨道的锥形束扫描不会收集完整的数据以进行体积[三维(3D)图像]的精确重建,因此可能会导致缺点或伪影。为了研究锥形束数据采集对图像质量的影响,我们评估了具有0.5毫米层厚的256层CT扫描仪原型的物理性能,并将其与具有0.75毫米层厚的16层CT扫描仪进行了比较。结果,我们发现图像噪声、均匀性和高对比度可检测性与z坐标无关。在失真测量中观察到费尔德坎普伪影。层厚灵敏度曲线(SSP)的半高宽(FWHM)随z坐标增加,尽管这似乎是由数据不完整以外的其他原因引起的。关于低对比度可检测性,较小的物体在中平面(z = 0毫米)比在z = 40毫米处检测得更清晰,尽管类似环形带的伪影会影响检测。16层和256层扫描仪之间的比较表明,在SSP、低对比度可检测性和失真方面,16层扫描仪性能更好。256层扫描仪在除失真测量(费尔德坎普伪影)之外的其他方面的劣势似乎部分是由扫描仪的原型性质导致的,应在未来的扫描仪中加以改进。两种CT的图像噪声、均匀性和高对比度可检测性几乎相同。256层扫描仪在点扩散函数(PSF)方面优于16层扫描仪,尽管这是由256层扫描仪较小的横向束宽导致的。为了在曝光剂量、噪声、层厚和横向空间分辨率方面全面比较两种扫描仪,计算了K = Dσ²h/a³,其中D是曝光剂量(CT剂量指数),σ是噪声幅度,h是层厚(SSP的FWHM),a是横向空间分辨率(PSF的FWHM)。结果表明,16层扫描仪的K值大25%,并且256层扫描仪在中平面的效率是16层扫描仪的1.25倍。256层扫描仪在K值方面的优势可能部分是由于广角锥形束扫描减少了浪费的曝光。尽管256层扫描仪存在几个问题,但如果分辨率由PSF或SSP的FWHM定义,在1秒的扫描中,它可以沿ζ轴在宽视野(约100毫米长)内获取横向分辨率约为1.0毫米、纵向分辨率约为1.3毫米的体积数据,这对于获取运动器官的动态3D(4D)图像应该非常有用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验