Development of real-time thermal neutron monitor using boron-loaded plastic scintillator with optical fiber for boron neutron capture therapy.

作者信息

Ishikawa M, Ono K, Sakurai Y, Unesaki H, Uritani A, Bengua G, Kobayashi T, Tanaka K, Kosako T

机构信息

Research Center for Nuclear Science and Technology, The University of Tokyo, 2-11-16 Yayoi Bunkyo-ku, Tokyo 113-0032, Japan.

出版信息

Appl Radiat Isot. 2004 Nov;61(5):775-9. doi: 10.1016/j.apradiso.2004.05.053.

Abstract

A new thermal neutron monitor for boron neutron capture therapy was developed in this study. We called this monitor equipped boron-loaded plastic scintillator that uses optical fiber for signal transmission as an [scintillator with optical fiber] SOF detector. A water phantom experiment was performed to verify how the SOF detector compared with conventional method of measuring thermal neutron fluence. Measurements with a single SOF detector yielded indistinguishable signals for thermal neutrons and gamma rays. To account for the gamma ray contribution in the signal recorded by the SOF detector, a paired SOF detector system was employed. This was composed of an SOF detector with boron-loaded scintillator and an SOF detector with a boron-free scintillator. The difference between the recorded counts of these paired SOF detectors was used as the measure of the gamma ray contribution in the measured neutron fluence. The paired SOF detectors were ascertained to be effective in measuring thermal neutron flux in the range above 10(6)(n/cm(2)/s). Clinical trials using paired SOF to measure thermal neutron flux during therapy confirmed that paired SOF detectors were effective as a real-time thermal neutron flux monitor.

摘要

本研究开发了一种用于硼中子俘获疗法的新型热中子监测器。我们将这种配备了含硼塑料闪烁体且使用光纤进行信号传输的监测器称为[带光纤的闪烁体]SOF探测器。进行了水模实验,以验证SOF探测器与测量热中子注量的传统方法相比如何。使用单个SOF探测器进行测量时,热中子和伽马射线产生的信号无法区分。为了考虑SOF探测器记录信号中的伽马射线贡献,采用了成对的SOF探测器系统。该系统由一个装有含硼闪烁体的SOF探测器和一个装有无硼闪烁体的SOF探测器组成。这两个成对的SOF探测器记录的计数之差用作测量中子注量中伽马射线贡献的量度。已确定成对的SOF探测器在测量高于10(6)(n/cm(2)/s)范围内的热中子通量时是有效的。使用成对SOF在治疗期间测量热中子通量的临床试验证实,成对的SOF探测器作为实时热中子通量监测器是有效的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验