Detachable coil for cerebral aneurysms: in vitro evaluation of magnetic field interactions, heating, and artifacts at 3T.

作者信息

Shellock Frank G, Gounis Matthew, Wakhloo Ajay

机构信息

Keck School of Medicine, University of Southern California and Institute for Magnetic Resonance Safety, Education, and Research, 7511 McConnell Avenue, Los Angeles, CA 90045, USA.

出版信息

AJNR Am J Neuroradiol. 2005 Feb;26(2):363-6.

Abstract

A detachable coil was assessed for magnetic field interactions, heating (body RF coil, SAR of 2.0-W/kg), and artifacts at 3-Tesla. The coil showed no magnetic field interactions and heating was negligible (+0.3 degrees C). Therefore, this coil is safe for a patient undergoing MR imaging at 3T or less. While artifacts may impair the ability to properly visualize anatomy in proximity to this implant, careful selection of imaging parameters can mitigate this problem.

摘要

相似文献

6
Drug eluting coronary stent: in vitro evaluation of magnet resonance safety at 3 Tesla.
J Cardiovasc Magn Reson. 2005;7(2):415-9. doi: 10.1081/jcmr-200053588.
8
Ventricular assist device implant (AB 5000) prototype cannula: in vitro assessment of MRI issues at 3-Tesla.
J Cardiovasc Magn Reson. 2008 May 21;10(1):23. doi: 10.1186/1532-429X-10-23.

引用本文的文献

1
Basic Concepts and Insights into Aortopulmonary Collateral Arteries in Congenital Heart Diseases.
Indian J Radiol Imaging. 2023 Jul 14;33(4):496-507. doi: 10.1055/s-0043-1770344. eCollection 2023 Oct.
2
Evaluation of magnetic resonance imaging issues for implantable microfabricated magnetic actuators.
Biomed Microdevices. 2014 Feb;16(1):153-61. doi: 10.1007/s10544-013-9815-3.
6
Radiofrequency-induced heating near fixed orthodontic appliances in high field MRI systems at 3.0 Tesla.
J Orofac Orthop. 2009 Nov;70(6):485-94. doi: 10.1007/s00056-009-9923-0. Epub 2009 Dec 4.
7
Body MRI artefacts: from image degradation to diagnostic utility.
Radiol Med. 2009 Feb;114(1):18-31. doi: 10.1007/s11547-008-0328-8. Epub 2008 Oct 4.
8
Ventricular assist device implant (AB 5000) prototype cannula: in vitro assessment of MRI issues at 3-Tesla.
J Cardiovasc Magn Reson. 2008 May 21;10(1):23. doi: 10.1186/1532-429X-10-23.
10
Modern cross-sectional imaging in the diagnosis and follow-up of intracranial aneurysms.
Eur Radiol. 2006 Sep;16(9):2051-66. doi: 10.1007/s00330-005-0092-2. Epub 2006 Jan 14.

本文引用的文献

2
Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system.
J Magn Reson Imaging. 2002 Dec;16(6):721-32. doi: 10.1002/jmri.10207.
4
MR Safety and the American College of Radiology White Paper.
AJR Am J Roentgenol. 2002 Jun;178(6):1349-52. doi: 10.2214/ajr.178.6.1781349.
7
Coronary arterial stents: safety and artifacts during MR imaging.
Radiology. 2000 Sep;216(3):781-7. doi: 10.1148/radiology.216.3.r00se03781.
10
Aneurysm clips: evaluation of MR imaging artifacts at 1.5 T.
Radiology. 1998 Nov;209(2):563-6. doi: 10.1148/radiology.209.2.9807590.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验