Suppr超能文献

基于7,16-二苄基-1,4,10,13-四氧杂-7,16-二氮杂环十八烷的电荷转移三碘离子选择性电极。

Charge-transfer triiodide ion-selective electrode based on 7,16-dibenzyl-1,4,10,13-tetraoxa-7,16-diazacyclooctadecane.

作者信息

Khayatian Golamreza, Rezatabar Hadi, Salimi Abdollah

机构信息

Department of Chemistry, Kurdistan University, Sanandaj, Iran.

出版信息

Anal Sci. 2005 Mar;21(3):297-302. doi: 10.2116/analsci.21.297.

Abstract

A new PVC membrane electrode for the triiodide ion based on a charge-transfer complex of iodine with 7,16-dibenzyl-1,4,10,13-tetraoxa-7,16-diazacyclooctadecane as a membrane carrier was prepared. The electrode exhibits a Nernstian response for triiodide ions over a wide concentration range (1.0 x 10(-1)-1.0 x 10(-5) M) with a slope of 59.3 +/- 0.9 mV decade(-1) and a detection limit of 6.3 x 10(-6) M. It has a response time of 30 s and can be used for at least 3 months without any divergence in the potential. The potentiometric response is independent of the pH, in the pH range 1.6 - 10.0. The proposed electrode has shown very high selectivity for the triiodide ion over a wide variety of other anions. This electrode was successfully applied as an indicator electrode in the potentiometric titration of ascorbic acid and hydroquinone from pharmaceutical preparations as well as ascorbic acid in orange juice and dissolved O2 in tap water.

摘要

制备了一种基于碘与7,16 - 二苄基 - 1,4,10,13 - 四氧杂 - 7,16 - 二氮杂环十八烷的电荷转移络合物的新型用于三碘离子的PVC膜电极。该电极在较宽的浓度范围(1.0×10⁻¹ - 1.0×10⁻⁵ M)内对三碘离子呈现能斯特响应,斜率为59.3±0.9 mV/decade⁻¹,检测限为6.3×10⁻⁶ M。其响应时间为30 s,可使用至少3个月而电位无任何偏差。电位响应在pH范围1.6 - 10.0内与pH无关。所提出的电极对三碘离子表现出对多种其他阴离子的极高选择性。该电极已成功用作指示电极,用于电位滴定药物制剂中的抗坏血酸和对苯二酚、橙汁中的抗坏血酸以及自来水中溶解的O₂。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验