Libicher M, Vetter M, Wolf I, Noeldge G, Kasperk C, Grafe I, Da Fonseca K, Hillmeier J, Meeder P J, Meinzer H P, Kauffmann G W
Department of Diagnostic Radiology, University of Heidelberg, Heidelberg, Germany.
Eur Radiol. 2005 Aug;15(8):1544-9. doi: 10.1007/s00330-005-2709-x. Epub 2005 Apr 5.
This study was intended to measure the volume of intravertebral cement after balloon kyphoplasty with high resolution computed tomography (CT) and dedicated software. Volume changes of biocompatible calcium phosphate cement (CPC) were detected during a follow-up of 12 months. Measurements were compared with a control group of patients treated with polymethylmethacrylate (PMMA). Twenty-three vertebrae (14 CPC, 9 PMMA) of 12 patients were examined with CT using an identical imaging protocol. Dedicated software was used to quantify intravertebral cement volume in subvoxel resolution by analyzing each cement implant with a density-weighted algorithm. The mean volume reduction of CPC was 0.08 ml after 12 months, which corresponds to an absorption rate of 2 vol%. However, the difference did not reach significance level (P>0.05). The mean error estimate was 0.005 ml, indicating excellent precision of the method. CT volumetry appears a precise tool for measurement of intravertebral cement volume. CT volumetry offers the possibility of in vivo measurement of CPC resorption.
本研究旨在通过高分辨率计算机断层扫描(CT)及专用软件测量球囊椎体后凸成形术后椎体内骨水泥的体积。在12个月的随访期间检测生物相容性磷酸钙骨水泥(CPC)的体积变化。将测量结果与接受聚甲基丙烯酸甲酯(PMMA)治疗的对照组患者进行比较。使用相同的成像方案对12例患者的23个椎体(14个CPC,9个PMMA)进行CT检查。通过密度加权算法分析每个骨水泥植入物,使用专用软件以亚体素分辨率量化椎体内骨水泥体积。12个月后CPC的平均体积减少0.08ml,对应吸收率为2%。然而,差异未达到显著水平(P>0.05)。平均误差估计为0.005ml,表明该方法具有出色的精度。CT容积测量似乎是测量椎体内骨水泥体积的精确工具。CT容积测量提供了在体内测量CPC吸收的可能性。