Suppr超能文献

LiF:Mg:Cu:P热释光剂量计的低能X射线响应:与LiF:Mg:Ti的比较。

The low energy X-ray response of the LiF:Mg:Cu:P thermoluminescent dosemeter: a comparison with LiF:Mg:Ti.

作者信息

Edwards C R, Mountford P J, Green S, Palethorpe J E, Moloney A J

机构信息

Medical Physics Directorate, University Hospital of North Staffordshire, Princes Road, Hartshill, Stoke-on-Trent, Staffordshire ST4 7LN, UK.

出版信息

Br J Radiol. 2005 Jun;78(930):543-7. doi: 10.1259/bjr/73133162.

Abstract

LiF:Mg:Cu:P thermoluminescent dosemeters (TLD) can be used for the same X-ray dosimetry applications as LiF:Mg:Ti, with each type having the disadvantage of a response dependent on energy, particularly at low energies. Measurements were made of the response per unit air kerma of LiF:Mg:Cu:P and LiF:Mg:Ti to nine quasi-monoenergetic X-ray beams with mean energies from 12 keV to 208 keV. Each measurement was normalized to the value produced by 6 MV X-rays. LiF:Mg:Cu:P was found to under-respond to a majority of these radiations whereas LiF:Mg:Ti over-responded to a majority. Their smallest relative measured response was produced by the lowest energy beam, and the maximum measured relative response of 1.15+/-0.07 and 1.21+/-0.07 for LiF:Mg:Cu:P and LiF:Mg:Ti, respectively, occurred at 33 keV. Energy response coefficients were derived from these measurements to estimate the error introduced by using either type of TLD to measure the dose from an X-ray spectrum different to that used for its absolute response calibration. It was calculated that if the response of either type of TLD was calibrated at 100 kVp, then an error of no more than +/-2% would be introduced into measurements of tube output at potentials of 50-130 kVp. LiF:Mg:Cu:P was found to introduce a larger error (up to 30%) into the measurement of body exit dose than LiF:Mg:Ti at tube potentials of 40-150 kVp, if its absolute response was calibrated using the corresponding body entrance beam. The method should allow this type of error to be estimated in other dosimetry applications for either type of TLD.

摘要

LiF:Mg:Cu:P热释光剂量计(TLD)可用于与LiF:Mg:Ti相同的X射线剂量测定应用,每种类型都有响应依赖于能量的缺点,特别是在低能量时。对LiF:Mg:Cu:P和LiF:Mg:Ti每单位空气比释动能对9束平均能量从12keV到208keV的准单能X射线束的响应进行了测量。每次测量都归一化为6MV X射线产生的值。发现LiF:Mg:Cu:P对这些辐射中的大多数响应不足,而LiF:Mg:Ti对大多数辐射响应过度。它们最小的相对测量响应是由最低能量束产生的,LiF:Mg:Cu:P和LiF:Mg:Ti分别在33keV时出现最大测量相对响应,分别为1.15±0.07和1.21±0.07。从这些测量中得出能量响应系数,以估计使用任何一种TLD测量与用于其绝对响应校准的X射线光谱不同的X射线光谱剂量时引入的误差。计算得出,如果任何一种TLD的响应在100kVp下校准,那么在50 - 130kVp的电位下测量管输出时引入的误差将不超过±2%。如果使用相应的身体入口束对LiF:Mg:Cu:P的绝对响应进行校准,发现在40 - 150kVp的管电位下,LiF:Mg:Cu:P在身体出射剂量测量中引入的误差(高达30%)比LiF:Mg:Ti更大。该方法应允许在其他剂量测定应用中对任何一种TLD估计这种类型的误差。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验