Coronary calcium measurements: effect of CT scanner type and calcium measure on rescan reproducibility--MESA study.

作者信息

Detrano Robert C, Anderson Melissa, Nelson Jennifer, Wong Nathan D, Carr J Jeffrey, McNitt-Gray Michael, Bild Diane E

机构信息

Div of Cardiology, Los Angeles Biomedical Research Inst, 1124 W Carson St, Bldg E-5, Torrance, CA 90502, USA.

出版信息

Radiology. 2005 Aug;236(2):477-84. doi: 10.1148/radiol.2362040513. Epub 2005 Jun 21.

Abstract

PURPOSE

To evaluate the effect of scanner type and calcium measure on the reproducibility of calcium measurements.

MATERIALS AND METHODS

This investigation was approved by the institutional review boards of each study site and by the Institutional Review Board of the Los Angeles Biomedical Research Institute. Informed consent for scanning and participation was obtained from all participants. The study was Health Insurance Portability and Accountability Act compliant. The Multi-Ethnic Study of Atherosclerosis (MESA) is a multicenter observational study of 6814 participants undergoing demographic, risk factor, and subclinical disease evaluations. Coronary artery calcium was measured by using duplicate CT scans. Three study centers used electron-beam computed tomography (CT), and three used multi-detector row CT. Coronary artery calcium was detected in 3355 participants. Three calcium measurement methods-Agatston score, calcium volume, and interpolated volume score-were evaluated. Mean absolute differences between calcium measures on scans 1 and 2, excluding cases for which both scans had a measure of zero, was modeled by using linear regression to compare reproducibility between scanner types. A repeated measures analysis of variance test was used to compare reproducibility across calcium measures, with mean percentage absolute difference as the outcome measure. Rescan reproducibility in relation to misregistrations, noise, and motion artifacts was also examined. Variables were log transformed to create a more normal distribution.

RESULTS

Concordance for presence of calcium between duplicate scans was high and similar for both electron-beam and multi-detector row CT (96%, kappa = 0.92). Mean absolute difference between calcium scores for the two scans was 15.8 for electron-beam and 16.9 for multi-detector row CT scanners (P = .06). Mean relative differences were 20.1 for Agatston score, 18.3 for calcium volume, and 18.3 for interpolated volume score (P < .01). Reproducibility was lower for scans with versus those without image misregistrations or motion artifacts (P < .01 for both).

CONCLUSION

Electron-beam and multi-detector row CT scanners have equivalent reproducibility for measuring coronary artery calcium. Calcium volumes and interpolated volume scores are slightly more reproducible than Agatston scores. Reproducibility is lower for scans with misregistrations or motion artifacts.

摘要

目的

评估扫描仪类型和钙测量方法对钙测量重复性的影响。

材料与方法

本研究经各研究地点的机构审查委员会以及洛杉矶生物医学研究所的机构审查委员会批准。所有参与者均获得了扫描和参与的知情同意书。该研究符合《健康保险流通与责任法案》。动脉粥样硬化多民族研究(MESA)是一项针对6814名参与者的多中心观察性研究,这些参与者接受了人口统计学、危险因素和亚临床疾病评估。使用重复CT扫描测量冠状动脉钙化。三个研究中心使用电子束计算机断层扫描(CT),三个使用多排探测器CT。在3355名参与者中检测到冠状动脉钙化。评估了三种钙测量方法——阿加斯顿积分、钙体积和插值体积积分。通过线性回归对扫描1和扫描2上钙测量值之间的平均绝对差异(不包括两次扫描测量值均为零的情况)进行建模,以比较不同扫描仪类型之间的重复性。使用重复测量方差分析检验来比较不同钙测量方法之间的重复性,以平均绝对差异百分比作为结果指标。还检查了与配准错误、噪声和运动伪影相关的重复扫描重复性。对变量进行对数转换以创建更正态的分布。

结果

重复扫描之间钙存在情况的一致性很高,电子束CT和多排探测器CT相似(96%,kappa = 0.92)。电子束CT两次扫描的钙积分平均绝对差异为15.8,多排探测器CT为16.9(P = 0.06)。阿加斯顿积分的平均相对差异为20.1,钙体积为18.3,插值体积积分为18.3(P < 0.01)。有图像配准错误或运动伪影的扫描重复性低于无此类情况的扫描(两者P < 0.01)。

结论

电子束CT和多排探测器CT扫描仪在测量冠状动脉钙化方面具有同等的重复性。钙体积和插值体积积分比阿加斯顿积分的重复性略高。有配准错误或运动伪影的扫描重复性较低。