Niwa Makoto, Watanabe Noriko, Ochiai Hisashi, Yamashita Kouwa
Pharmacokinetics Group, Technical Development Department, Research and Development Division, Pharmaceuticals Group, Nippon Kayaku Co. Ltd., 3-31-12, Shimo, Kita-ku, Tokyo 115-8588, Japan.
J Chromatogr B Analyt Technol Biomed Life Sci. 2005 Sep 25;824(1-2):258-66. doi: 10.1016/j.jchromb.2005.07.033.
A quantitative method for measuring testosterone (T) concentrations in rat plasma was developed using ethyl oxime and acetyl ester derivatization and liquid chromatography-atmosphere pressure chemical ionization tandem mass spectrometry (LC-APCI-MS/MS). The method utilizes a solid phase extraction with Varian Bond Elut C18, a derivatization process to form testosterone ethoxime acetate and LC-APCI-MS/MS with a reversed phase LC and a C8 column. This method is capable of detecting testosterone concentrations as low as 0.2 ng/ml in a 0.05 ml sample of rat plasma. This method can be used as a sensitive chromatography-based assay for small sample volumes of rat blood.
采用乙肟和乙酰酯衍生化以及液相色谱-大气压化学电离串联质谱法(LC-APCI-MS/MS),开发了一种测定大鼠血浆中睾酮(T)浓度的定量方法。该方法采用瓦里安Bond Elut C18进行固相萃取,通过衍生化过程形成睾酮乙氧肟酸酯,并利用反相液相色谱和C8柱进行LC-APCI-MS/MS分析。该方法能够在0.05 ml大鼠血浆样本中检测低至0.2 ng/ml的睾酮浓度。该方法可作为一种灵敏的基于色谱的大鼠小体积血液检测方法。