Pre-clinical validation of a new intra-operative "dual beam Doppler" blood flowmeter in an artificial circuit.

作者信息

Cikirikcioglu Mustafa, Cikirikcioglu Y Banu, Khabiri Ebrahim, Djebaili M Karim, Kalangos Afksendiyos, Walpoth Beat H

机构信息

Department of Cardiovascular Surgery, University Hospital of Geneva, Geneva, Switzerland.

出版信息

Heart Surg Forum. 2006;9(1):E499-505; discussion E505. doi: 10.1532/HSF98.20051031.

Abstract

BACKGROUND

Intra-operative flow measurement during coronary or peripheral bypass operations is helpful for ruling out technical failures and for prediction of complication and patency rates. Preclinical validation of the flowmeters is required in order to rely on the intra-operatively measured results. The aim of this study is to evaluate a new "dual beam Doppler" blood flowmeter before clinical application and to compare it with the established "transit time flow measure-ment" technique in an artificial circuit.

METHODS

Measurements were performed in an experimental flow model using pig blood and pig arteries. Three different flowmeters were used: Quantix OR (dual beam doppler flowmeter), CardioMed (transit time flowmeter), and Transonic (transit time flowmeter). Three validation tests were performed to assess correlation, precision, and repeatability of devices. (1) Correlation and agreement analysis was performed with various flow amounts (10-350 mL/min) (n = 160). (2) Device reproducibility and measurement stability were tested with a constant flow (flow amount = 300 mL/min) (n = 30). (3) A user accuracy test (intra- and inter-observer variability) was performed by 5 different observers with a constant flow (flow amount = 205 mL/min) (n = 75). Time collected true flow was used as a reference method in all steps and all tests were performed in a blind manner. Results are shown as mean values +/- standard deviations. Pear-son's correlation and Bland-Altman plot analyses were used to compare measurements.

RESULTS

The mean flow was 167 +/- 98 mL/min for true flow and 162 +/- 94 mL/min, 165 +/- 94 mL/min, and 166 +/- 100 mL/min for Quantix OR, CardioMed, and Transonic, respectively. Correlation coefficients between Quantix OR, Medi-Stim, Transonic, and time collected true flow were over 0.98 (P = .01). Most of the measured results ( > 90%) were between +/- 1.96 SD agreement limits in Bland and Altman plot analysis. All devices showed good results in the reproducibility test. During the user accuracy test, larger variance changes were observed between intra- and inter-observer results with the dual beam Doppler flowmeter compared to the 2 used transit time flowmeters when used for single sided vessel access without stabilization device (available from the manufacturer).

CONCLUSION

All 3 tested flowmeters showed an excellent correlation to the true flow in an artificial circuit and the accuracy of the tested devices was within agreement limits. Reproducibility of all devices was good and linear. The new dual beam Doppler flow measurement technique compares favorably to the classic transit time method. Clinical use may depend on operator, location, and condition, thus more studies may be required to ensure uniform results using the currently available blood flow measurement devices.

摘要

背景

在冠状动脉或外周血管搭桥手术中进行术中血流测量,有助于排除技术故障,并预测并发症和通畅率。为了信赖术中测量结果,需要对流量计进行临床前验证。本研究的目的是在临床应用前评估一种新型“双束多普勒”血流计,并在人工循环中与已确立的“渡越时间血流测量”技术进行比较。

方法

在使用猪血和猪动脉的实验血流模型中进行测量。使用了三种不同的流量计:Quantix OR(双束多普勒流量计)、CardioMed(渡越时间流量计)和Transonic(渡越时间流量计)。进行了三项验证测试,以评估设备的相关性、精度和可重复性。(1)对各种流量(10 - 350 mL/分钟)进行相关性和一致性分析(n = 160)。(2)在恒定流量(流量 = 300 mL/分钟)下测试设备的可重复性和测量稳定性(n = 30)。(3)由5名不同的观察者在恒定流量(流量 = 205 mL/分钟)下进行用户准确性测试(观察者内和观察者间的变异性)(n = 75)。在所有步骤中,将采集到的真实血流时间用作参考方法,所有测试均采用盲法进行。结果以平均值±标准差表示。使用Pearson相关性分析和Bland - Altman图分析来比较测量结果。

结果

真实血流的平均流量为167±98 mL/分钟,Quantix OR、CardioMed和Transonic的平均流量分别为162±94 mL/分钟、165±94 mL/分钟和166±100 mL/分钟。Quantix OR、Medi - Stim、Transonic与采集到的真实血流时间之间的相关系数均超过0.98(P = 0.01)。在Bland和Altman图分析中,大多数测量结果(> 90%)在±1.96标准差一致性范围内。所有设备在可重复性测试中均显示出良好的结果。在用户准确性测试中,与两种使用的渡越时间流量计相比,当用于没有稳定装置(制造商提供)的单侧血管接入时,双束多普勒流量计在观察者内和观察者间结果之间观察到更大的方差变化。

结论

所有三种测试的流量计在人工循环中与真实血流均显示出极好的相关性,测试设备的准确性在一致性范围内。所有设备的可重复性良好且呈线性。新型双束多普勒血流测量技术与经典的渡越时间方法相比具有优势。临床应用可能取决于操作者、位置和条件,因此可能需要更多研究以确保使用当前可用的血流测量设备能获得一致的结果。