Stickler D J, Jones S M, Adusei G O, Waters M G
Cardiff School of Biosciences, Cardiff University, Cardiff CF10 3TL, Wales, United Kingdom.
J Clin Microbiol. 2006 Apr;44(4):1540-2. doi: 10.1128/JCM.44.4.1540-1542.2006.
A simple sensor has been developed to detect the early stages of urinary catheter encrustation and avoid the clinical crises induced by catheter blockage. In laboratory models of colonization by Proteus mirabilis, the sensor signaled encrustation at an average time of 43 h before catheters were blocked with crystalline biofilm.
一种简单的传感器已被开发出来,用于检测导尿管结壳的早期阶段,并避免由导管堵塞引发的临床危机。在奇异变形杆菌定植的实验室模型中,该传感器在导管被结晶生物膜堵塞前平均43小时发出结壳信号。