Suppr超能文献

New flowmetric measurement methods of power dissipated by an ultrasonic generator in an aqueous medium.

作者信息

Mancier Valérie, Leclercq Didier

机构信息

D.T.I., E.A. 3083, UFR Sciences Exactes et Naturelles, BP 1039, F-51687 Reims Cedex 2, France.

出版信息

Ultrason Sonochem. 2007 Feb;14(2):99-106. doi: 10.1016/j.ultsonch.2006.03.004. Epub 2006 May 3.

Abstract

Two new determination methods of the power dissipated in an aqueous medium by an ultrasound generator were developed. They are based on the use of a heat flow sensor inserted between a tank and a heat sink that allows to measure the power directly coming through the sensor. To be exploitable, the first method requires waiting for stationary flow. On the other hand, the second, extrapolated from the first one, makes it possible to determine the dissipated power in only five minutes. Finally, the results obtained with the flowmetric method are compared to the classical calorimetric ones.