Kondepati Venkata Radhakrishna, Damm Uwe, Heise H Michael
ISAS-Institute for Analytical Sciences at the University of Dortmund, Bunsen-Kirchhoff-Strasse 11, D-44139 Dortmund, Germany.
Appl Spectrosc. 2006 Aug;60(8):920-5. doi: 10.1366/000370206778062066.
Application of mid-infrared spectroscopy for the determination of urea in blood plasma dialysates of microliter sample volumes using a transmission microcell was investigated. Infrared spectra of the dialysates of plasma samples collected from 75 different patients using CMA 60 microdialysis catheters were evaluated with multivariate partial least squares regression. Using the absorbance spectral data from 1520-1420 cm(-1) and 1220-1120 cm(-1), a minimum standard error of prediction (SEP) of 0.88 mg/dL (0.14 mM) was achieved with spectral variable selection. Our findings suggest the feasibility of developing a mid-infrared sensor in combination with micro-fluidics for on-line monitoring of urea in patients undergoing dialysis treatment.
研究了使用透射微池,将中红外光谱法应用于测定微升样品体积的血浆透析液中的尿素。使用CMA 60微透析导管收集的75名不同患者的血浆样品透析液的红外光谱,通过多元偏最小二乘回归进行评估。利用1520 - 1420 cm(-1)和1220 - 1120 cm(-1)的吸光光谱数据,通过光谱变量选择实现了0.88 mg/dL(0.14 mM)的最小预测标准误差(SEP)。我们的研究结果表明,结合微流体技术开发用于透析治疗患者尿素在线监测的中红外传感器是可行的。