高强度聚焦超声治疗子宫肌瘤的疗效评估:超声造影的应用价值

Therapeutic response assessment of high intensity focused ultrasound therapy for uterine fibroid: utility of contrast-enhanced ultrasonography.

作者信息

Zhou Xiao Dong, Ren Xiao Long, Zhang Jun, He Guang Bin, Zheng Min Juan, Tian Xue, Li Li, Zhu Ting, Zhang Min, Wang Lei, Luo Wen

机构信息

Department of Ultrasonography, Xijing Hospital, The Fourth Military Medical University, No. 17 West Changle Road, Xi'an 710032, China.

出版信息

Eur J Radiol. 2007 May;62(2):289-94. doi: 10.1016/j.ejrad.2006.11.039. Epub 2007 Jan 26.

Abstract

PURPOSE

To assess the utility of contrast-enhanced ultrasonography (ceUS) in the assessment of the therapeutic response to high intensity focused ultrasound (HIFU) ablation in patients with uterine fibroid.

MATERIALS AND METHODS

Sixty-four patients with a total of 64 uterine fibroids (mean: 5.3+/-1.2 cm; range: 3.2-8.9 cm) treated with HIFU ablation under the ultrasound guidance were evaluated with ceUS after receiving an intravenous bolus injection of a microbubble contrast agent (SonoVue) within 1 week after intervention. We obtained serial ceUS images during the time period from beginning to 5 min after the initiation of the bolus contrast injection. All of the patients underwent a contrast enhanced MRI (ceMRI) and ultrasound guided needle puncture biopsy within 1 week after HIFU ablation. And as a follow-up, all of the patients underwent US at 1, 3, 6 and 12 months after HIFU treatment. The volume change was observed and compared to pre- and post-HIFU ablation. The results of the ceUS were compared with those of the ceMRI in terms of the presence or absence of residual unablated tumor and pathologic change in the treated lesions.

RESULTS

On ceUS, diagnostic accuracy was 100%, while residual unablated tumors were found in three uterine fibroids (4.7%) and failed treatment was found in eight uterine fibroids (12.5%). All the 11 fibroids were subjected to additional HIFU ablation. Of the 58 ablated fibroids without residual tumors on both the ceUS and ceMRI after the HIFU ablation, the volumes of all the fibroids decreased in different degrees during the 1 year follow-up USs. And histologic examinations confirmed findings of necrotic and viable tumor tissue, respectively.

CONCLUSION

CEUS is potentially useful for evaluating the early therapeutic effect of percutaneous HIFU ablation for uterine fibroids.

摘要

目的

评估超声造影(ceUS)在评估子宫肌瘤患者高强度聚焦超声(HIFU)消融治疗反应中的应用价值。

材料与方法

64例共64个子宫肌瘤(平均大小:5.3±1.2 cm;范围:3.2 - 8.9 cm)患者在超声引导下接受HIFU消融治疗,干预后1周内静脉推注微泡造影剂(声诺维)后行ceUS评估。在推注造影剂开始至5分钟内获取系列ceUS图像。所有患者在HIFU消融后1周内接受对比增强磁共振成像(ceMRI)和超声引导下穿刺活检。作为随访,所有患者在HIFU治疗后1、3、6和12个月接受超声检查。观察体积变化并与HIFU消融前后进行比较。将ceUS结果与ceMRI结果在残留未消融肿瘤的有无及治疗病变的病理变化方面进行比较。

结果

ceUS诊断准确率为100%,3个子宫肌瘤(4.7%)发现残留未消融肿瘤,8个子宫肌瘤(12.5%)发现治疗失败。所有这11个肌瘤均接受了额外的HIFU消融。在HIFU消融后ceUS和ceMRI均未发现残留肿瘤的58个消融肌瘤中,所有肌瘤在1年随访超声检查期间体积均有不同程度减小。组织学检查分别证实了坏死和存活肿瘤组织的情况。

结论

CEUS对评估经皮HIFU消融子宫肌瘤的早期治疗效果可能有用。