Reuland Mirjam S, Reuland Andreas J, Nishi Yutaro, Auffarth Gerd U
Heidelberg Intraocular Lens and Refractive Surgery Research Group, Heidelberg, Germany.
J Refract Surg. 2007 Apr;23(4):368-73. doi: 10.3928/1081-597X-20070401-09.
Corneal radii (R1, R2) and anterior chamber depth are important parameters for biometry and refractive surgery. This study aimed to investigate whether the IOLMaster, a biometric device, and the Pentacam, a rotating Scheimpflug camera, provide comparable results.
In this prospective study, corneal radii and anterior chamber depth were analyzed in 82 eyes of 41 phakic patients (median age 72 years) using the IOLMaster and Pentacam. Normal distribution was confirmed using the Kolmogorov-Smirnov test. A paired t test was used for statistical analysis.
No statistically significant difference was noted for R2 using the Pentacam versus the IOLMaster (P = .40). There was a small statistical difference of 0.03 mm (P < .01) for R1 (corresponding to 0.08 diopters [D]). The IOLMaster measured a mean R1 of 7.87 mm and R2 of 7.67 mm. The Pentacam measured mean R1 of 7.90 mm and R2 of 7.69 mm. For anterior chamber depth values, a small statistically significant difference of 0.05 mm (P < .001) was found (corresponding to 0.01 D). Mean anterior chamber depth measured from the epithelium was 3.20 mm for the IOLMaster and 3.25 mm for the Pentacam. The Bland-Altman plot showed no distorting trends for either variable.
Keratometry and anterior chamber depth were comparable for the two devices. If the axial length is known, the Pentacam can be used as a biometric device.
角膜半径(R1、R2)和前房深度是生物测量和屈光手术的重要参数。本研究旨在调查生物测量设备IOLMaster和旋转式Scheimpflug相机Pentacam是否能提供可比的结果。
在这项前瞻性研究中,使用IOLMaster和Pentacam对41例有晶状体患者(中位年龄72岁)的82只眼睛进行角膜半径和前房深度分析。使用Kolmogorov-Smirnov检验确认正态分布。采用配对t检验进行统计分析。
Pentacam与IOLMaster测量的R2无统计学显著差异(P = 0.40)。R1存在0.03 mm的微小统计学差异(P < 0.01)(相当于0.08屈光度[D])。IOLMaster测量的R1平均值为7.87 mm,R2为7.67 mm。Pentacam测量的R1平均值为7.90 mm,R2为7.69 mm。对于前房深度值,发现存在0.05 mm的微小统计学显著差异(P < 0.001)(相当于0.01 D)。IOLMaster测量的从上皮起的平均前房深度为3.20 mm,Pentacam为3.25 mm。Bland-Altman图显示两个变量均无扭曲趋势。
两种设备的角膜曲率测量和前房深度具有可比性。如果已知眼轴长度,Pentacam可作为一种生物测量设备使用。