Europace. 2007 Aug;9(8):687-93. doi: 10.1093/europace/eum066. Epub 2007 May 3.
We examine an expert system designed to permanently monitor patients with congestive heart failure (CHF) using data of a dual-sensor pacemaker and to allow warning of significant changes in physiological indices.
This study included 67 implanted patients divided into two groups: a control group without history of CHF (n = 19) who had received DDDR pacemakers (DDD group) and a test group (n = 48) who had received cardiac resynchronization therapy systems (CRT group) for severe CHF (NYHA III or IV, LVEF <40%). The embedded monitoring system measures minute ventilation (MV) and activity (ACT) at rest and at exercise. All devices collect data, and all adverse medical events were recorded. Data are stored daily for up to 3 months. The mean ACT was similar for both groups. Mean rest and exercise MV were significantly higher in CRT group. On 195 periods of 1-month follow-up in the CRT group, 31 events were suspected, 22 were true positive, 9 were false-positive, and 3 clinical adverse events were not predicted (sensitivity: 88%, specificity: 94.7%, positive predictive value: 71%, negative predictive value: 98.2%)
A new diagnostic expert system that holds promise for the long-term ambulatory monitoring of CHF was developed.
我们研究了一种专家系统,该系统旨在利用双传感器起搏器的数据对充血性心力衰竭(CHF)患者进行长期监测,并对生理指标的显著变化发出警报。
本研究纳入了67例植入患者,分为两组:无CHF病史的对照组(n = 19),他们接受了DDDR起搏器(DDD组);以及测试组(n = 48),他们因严重CHF(纽约心脏协会III或IV级,左心室射血分数<40%)接受了心脏再同步治疗系统(CRT组)。嵌入式监测系统在静息和运动时测量分钟通气量(MV)和活动量(ACT)。所有设备收集数据,并记录所有不良医疗事件。数据每天存储,最长存储3个月。两组的平均ACT相似。CRT组的静息和运动时平均MV显著更高。在CRT组的195个1个月随访期内,怀疑有31起事件,其中22起为真阳性,9起为假阳性,3起临床不良事件未被预测(敏感性:88%,特异性:94.7%,阳性预测值:71%,阴性预测值:98.2%)
开发了一种有望用于CHF长期门诊监测的新型诊断专家系统。