使用高强度聚焦超声切除腹部实体器官。

Resection of abdominal solid organs using high-intensity focused ultrasound.

作者信息

Zderic Vesna, O'Keefe Grant E, Foley Jessica L, Vaezy Shahram

机构信息

Department of Bioengineering, University of Washington, Seattle, WA 98195, USA.

出版信息

Ultrasound Med Biol. 2007 Aug;33(8):1251-8. doi: 10.1016/j.ultrasmedbio.2007.02.010. Epub 2007 May 11.

Abstract

Our objective was to evaluate high-intensity focused ultrasound (HIFU) for minimizing blood loss during surgery by hemodynamically isolating large portions of solid organs before their resection. A high-power HIFU device (in situ intensity of 9000 W/cm(2), frequency of 3.3 MHz) was used to produce a wall of cautery for sealing of blood vessels along the resection line in surgically exposed solid organs (liver lobes, spleen and kidneys) of eight adult pigs. Following HIFU application, the distal portion of the organ was excised using a scalpel. If any blood vessels were still bleeding, additional HIFU application was used to stop the bleeding. The resection was achieved in 6.0 +/- 1.5 min (liver), 3.6 +/- 1.1 min (spleen) and 2.8 +/- 0.6 min (kidneys) of HIFU treatment time, with no occurrence of bleeding for up to 4 h (until sacrifice). The coagulated region at the resection line had average width of 3 cm and extended through the whole thickness of the organ (up to 4 cm). Blood vessels of up to 1 cm in size were occluded. This method holds promise for future clinical applications in resection of solid tumors and hemorrhage control from high-grade organ injuries.

摘要

我们的目标是评估高强度聚焦超声(HIFU)在手术中通过在切除实体器官的大部分组织前进行血流动力学隔离来减少失血的效果。使用高功率HIFU设备(原位强度为9000 W/cm²,频率为3.3 MHz)在8只成年猪的手术暴露的实体器官(肝叶、脾脏和肾脏)中沿着切除线产生一道烧灼壁以封闭血管。在应用HIFU后,使用手术刀切除器官的远端部分。如果有任何血管仍在出血,则再次应用HIFU来止血。肝、脾、肾的HIFU治疗时间分别为6.0±1.5分钟、3.6±1.1分钟和2.8±0.6分钟,切除过程中无出血,直至4小时后(直至处死后)。切除线处的凝固区域平均宽度为3 cm,贯穿器官全层(达4 cm)。直径达1 cm的血管被阻塞。该方法在实体肿瘤切除和控制高级别器官损伤出血的未来临床应用中具有前景。

相似文献

1
Resection of abdominal solid organs using high-intensity focused ultrasound.
Ultrasound Med Biol. 2007 Aug;33(8):1251-8. doi: 10.1016/j.ultrasmedbio.2007.02.010. Epub 2007 May 11.
2
Use of high-intensity focused ultrasound to control bleeding.
J Vasc Surg. 1999 Mar;29(3):533-42. doi: 10.1016/s0741-5214(99)70282-x.
3
High-intensity focused ultrasound (HIFU)-assisted hepatic resection in an animal model.
Ann Surg Oncol. 2012 Jul;19 Suppl 3:S447-54. doi: 10.1245/s10434-011-1875-0. Epub 2011 Jul 28.
5
Control of splenic bleeding by using high intensity ultrasound.
J Trauma. 1999 Sep;47(3):521-5. doi: 10.1097/00005373-199909000-00015.
6
Hemostasis and sealing of air leaks in the lung using high-intensity focused ultrasound.
J Trauma. 2007 Jun;62(6):1390-5. doi: 10.1097/01.ta.0000215942.42423.6c.
7
Hemostasis of punctured vessels using Doppler-guided high-intensity ultrasound.
Ultrasound Med Biol. 1999 Jul;25(6):985-90. doi: 10.1016/s0301-5629(99)00027-7.
8
Pulsatile flow phantom for ultrasound image-guided HIFU treatment of vascular injuries.
Ultrasound Med Biol. 2007 Aug;33(8):1269-76. doi: 10.1016/j.ultrasmedbio.2007.02.011. Epub 2007 Apr 27.
9
Hemorrhage control in arteries using high-intensity focused ultrasound: a survival study.
Ultrasonics. 2006 Jan;44(1):46-53. doi: 10.1016/j.ultras.2005.08.002. Epub 2005 Sep 2.
10
Hemostasis using high intensity focused ultrasound.
Eur J Ultrasound. 1999 Mar;9(1):79-87. doi: 10.1016/s0929-8266(99)00014-2.

引用本文的文献

1
Image-guided ultrasound phased arrays are a disruptive technology for non-invasive therapy.
Phys Med Biol. 2016 Sep 7;61(17):R206-48. doi: 10.1088/0031-9155/61/17/R206. Epub 2016 Aug 5.
3
High-intensity focused ultrasound ablation: an effective and safe treatment for secondary hypersplenism.
Br J Radiol. 2014 Nov;87(1043):20140374. doi: 10.1259/bjr.20140374. Epub 2014 Aug 20.
4
The Nd:YAG LIMAX® 120 high-output laser: local effects and resection capacity on liver parenchyma.
Lasers Med Sci. 2014 Jul;29(4):1411-6. doi: 10.1007/s10103-014-1544-1. Epub 2014 Feb 22.

本文引用的文献

1
High intensity ultrasound clamp for bloodless partial nephrectomy: In vitro and in vivo experiments.
Ultrasound Med Biol. 2007 Jan;33(1):105-12. doi: 10.1016/j.ultrasmedbio.2006.07.028.
2
Haemostatic partial nephrectomy using bipolar radiofrequency ablation.
BJU Int. 2005 Nov;96(7):1101-4. doi: 10.1111/j.1464-410X.2005.05743.x.
3
Clinical outcome of 214 liver resections using microwave tissue coagulation.
Hepatogastroenterology. 2005 Jul-Aug;52(64):1180-5.
4
Controlled study of in-line ovine spleen transection assisted by radiofrequency ablation.
J Trauma. 2005 Apr;58(4):841-4. doi: 10.1097/01.ta.0000161495.29886.24.
7
Liver resection using a saline-linked radiofrequency dissecting sealer for transection of the liver.
J Am Coll Surg. 2005 Feb;200(2):308-13. doi: 10.1016/j.jamcollsurg.2004.10.008.
8
High-intensity ultrasound treatment of blunt abdominal solid organ injury: an animal model.
J Trauma. 2004 Jul;57(1):152-6. doi: 10.1097/01.ta.0000088009.93307.23.
9
Multiple gene differential expression patterns in human ischemic liver: safe limit of warm ischemic time.
World J Gastroenterol. 2004 Jul 15;10(14):2130-3. doi: 10.3748/wjg.v10.i14.2130.
10
Liver hemostasis with high-intensity ultrasound: repair and healing.
J Ultrasound Med. 2004 Feb;23(2):217-25. doi: 10.7863/jum.2004.23.2.217.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验