Yoshida H, Kakino T, Kajitani M, Goh K, Gohda T, Yasuda K, Tanabe T
Cardiovascular Surgery, Hokkaido University, Sapporo, Japan.
ASAIO Trans. 1991 Jul-Sep;37(3):M272-3.
A new aortic prosthesis (SAP, shape memory aortic prosthesis) was developed with its transfemoral placement system to treat aortic dissection with minimal invasion. The SAP consists of a Nitinol stent and polyurethane tube, and it is designed in such a configuration that it can be compressed inside a 14 Fr catheter when cold. It regains its original shape when warmed to 30 degrees C. From the results of intra-aortic sutureless placement of SAP in 10 mongrel dogs, we concluded that a SAP of proper diameter could be implanted safely for more than 3 months. The transfemoral placement system is composed of a delivery catheter, pushing rod, catching catheter, and pulling wire. Using these devices, transfemoral placement of SAP was demonstrated successfully in three dogs and one sheep. The original SAP and delivery system can be used as an emergency procedure for aortic dissections, such as Stanford type B. The advantages of the procedure are minimal invasion of acutely ill patients and prompt application in the catheter laboratory.
一种新型主动脉假体(形状记忆主动脉假体,SAP)及其经股动脉置入系统被研发出来,用于以微创方式治疗主动脉夹层。SAP由镍钛诺支架和聚氨酯管组成,其设计结构使其在低温时能够被压缩进一根14F的导管内。当加热到30摄氏度时,它会恢复到原来的形状。根据在10只杂种犬体内进行的SAP主动脉内无缝置入的结果,我们得出结论,合适直径的SAP能够安全植入超过3个月。经股动脉置入系统由输送导管、推杆、抓捕导管和牵引线组成。利用这些装置,在3只犬和1只羊身上成功演示了SAP的经股动脉置入。最初的SAP和输送系统可作为主动脉夹层(如斯坦福B型)的紧急治疗手段。该手术的优点是对危重病患者的侵入性最小,且能在导管室迅速实施。