Thompson Matthew S, Cui Weidong, Reilly James P
Department of Chemistry, Indiana University, Bloomington, Indiana 47405-4001, USA.
J Am Soc Mass Spectrom. 2007 Aug;18(8):1439-52. doi: 10.1016/j.jasms.2007.04.015. Epub 2007 Apr 29.
Several groups have investigated the photodissociation of peptide ions with ultraviolet light. Significant differences have been reported with 157 and 193 nm excitation. Recent studies have shown that the mass analyzer can also influence the observed photofragment distribution. Comparison of experiments using different peptides, wavelengths, and mass analyzers is undesirably complicated. In the present work, several peptides are analyzed with both 157 and 193 nm photodissociation in tandem-TOF and linear ion trap mass spectrometers. The results indicate that the fragment ion distribution can be influenced by both the photodissociation wavelength and the mass analyzer. The two wavelengths generate similar spectra in an ion trap but quite different results in a tandem-TOF instrument.
多个研究小组对肽离子在紫外光下的光解离进行了研究。已有报道称,157纳米和193纳米激发下存在显著差异。最近的研究表明,质量分析器也会影响所观察到的光碎片分布。使用不同肽、波长和质量分析器的实验比较复杂,令人难以处理。在本研究中,使用串联飞行时间质谱仪和线性离子阱质谱仪,对几种肽分别进行了157纳米和193纳米光解离分析。结果表明,碎片离子分布会受到光解离波长和质量分析器的影响。这两个波长在离子阱中产生相似的光谱,但在串联飞行时间仪器中产生的结果却大不相同。