使用经阴道三维超声和多普勒血管造影术评估孕期宫颈体积和血流指数时观察者间及观察者自身的一致性。

Intra- and interobserver agreement on cervical volume and flow indices during pregnancy using transvaginal 3-dimensional ultrasonography and Doppler angiography.

作者信息

Basgul Alin, Kavak Zehra N, Bakirci Nadi, Gokaslan Husnu

机构信息

Marmara University School of Medicine, Department of Obstetrics and Gynecology, Fetal Medicine Unit, Istanbul, Turkey.

出版信息

Int J Fertil Womens Med. 2006 Nov-Dec;51(6):256-61.

Abstract

OBJECTIVES

To assess intra- and interobserver agreement in cervical volume and flow indices measurements.

METHOD

We prospectively examined 126 patients by two seperate observers using transvaginal 3D gray-scale and power Doppler ultrasound. The two acquired volume datasets were analyzed using the VOCAL imaging program for assessing cervical volume, vascularization index (VI), flow index (FI), and vascularization flow index (VFI). Reproducibility of volume and vascularity measurement was assessed by calculating intraclass (intra-CC) and interclass (inter-CC) correlation coefficients (ICCs).

RESULTS

Both intraobserver and interobserver cervical volume measurements were in perfect agreement with intra-CC values of 0.95, 0.96 for both examiners and with an inter-CC value of 0.95. Intraobserver agreement for VI, FI and VFI measurements were as good as the interobserver agreement for VI, and VFI measurements were adequate but less for FI measurements (inter-CC 0.67). Overall, volumetric data were more reliably acquirable than power Doppler measurements.

CONCLUSIONS

3D ultrasound gray-scale and power Doppler measurement of cervical volume and vascularization have acceptable intra- and interobserver variations and thus may be used in clinical research of cervical physiology and pathophysiology during pregnancy.

摘要

目的

评估观察者内和观察者间在宫颈体积和血流指数测量方面的一致性。

方法

我们由两名独立观察者对126例患者进行前瞻性经阴道三维灰阶和能量多普勒超声检查。使用VOCAL成像程序分析采集的两个体积数据集,以评估宫颈体积、血管化指数(VI)、血流指数(FI)和血管化血流指数(VFI)。通过计算组内(intra-CC)和组间(inter-CC)相关系数(ICC)来评估体积和血管测量的可重复性。

结果

观察者内和观察者间的宫颈体积测量结果均高度一致,两位检查者的组内相关系数值分别为0.95、0.96,组间相关系数值为0.95。VI、FI和VFI测量的观察者内一致性与VI的观察者间一致性相当,VFI测量的观察者间一致性尚可,但FI测量的观察者间一致性较差(组间相关系数为0.67)。总体而言,体积数据比能量多普勒测量更可靠。

结论

经阴道三维超声灰阶和能量多普勒对宫颈体积和血管化的测量具有可接受的观察者内和观察者间差异,因此可用于孕期宫颈生理和病理生理的临床研究。