Evaluation of Verisyse and Artiflex phakic intraocular lenses during accommodation using Visante optical coherence tomography.

作者信息

Güell José Luis, Morral Merce, Gris Oscar, Gaytan Javier, Sisquella Maite, Manero Felicidad

机构信息

Instituto de Microcirugia Ocular, Barcelona, Spain.

出版信息

J Cataract Refract Surg. 2007 Aug;33(8):1398-404. doi: 10.1016/j.jcrs.2007.04.026.

Abstract

PURPOSE

To perform a dynamic study of the relationship between Verisyse (AMO) and Artiflex (Ophtec B.V.) phakic intraocular lenses (pIOLs) and anterior chamber structures during accommodation using optical coherence tomography (OCT) (Visante, Carl Zeiss Meditec, Inc.)

SETTING

Institutional practice.

METHODS

Eleven myopic patients were randomly selected to have implantation of a Verisyse pIOL in 1 eye and an Artiflex pIOL in the other. Using a 2-dimensional image, dynamic measurements of the relationship between the anterior surface of the pIOL and the corneal endothelium, the posterior surface of the pIOL and the anterior surface of the crystalline lens, and the pupil diameter were performed using Visante OCT. Physiological accommodation was stimulated by adding lenses in 1.00 diopter (D) steps from +1.00 to -7.00 D.

RESULTS

Both groups had a significant decrease in pupil diameter (P<.0001, generalized linear model [GLM]) and in the distance between the anterior surface of the pIOL and the corneal endothelium (P<.0001, GLM) with accommodation. There were no statistically significant changes in the distance between the posterior surface of either pIOL and the anterior surface of the crystalline lens (P = .2845, GLM). There were no statistically significant differences between the 2 pIOLs in any measurement (P>.05, GLM).

CONCLUSIONS

The results fit with Helmholtz' theory of accommodation as forward movement of the diaphragm iris-crystalline lens was seen. There was a decrease in the distance between the pIOL and corneal endothelium and in the pupil diameter, whereas the distance between both pIOLs and the crystalline lens remained constant throughout the accommodation examination. This suggests that the risk for cataract from intermittent contact between the crystalline lens and IOL from accommodative effort is unlikely.

摘要

目的

使用光学相干断层扫描(OCT)(Visante,卡尔蔡司医疗技术公司)对Verisyse(AMO)和Artiflex(Ophtec B.V.)有晶状体眼人工晶状体(pIOL)与前房结构在调节过程中的关系进行动态研究。

设置

机构实践。

方法

随机选择11例近视患者,一只眼植入Verisyse pIOL,另一只眼植入Artiflex pIOL。使用二维图像,通过Visante OCT对pIOL前表面与角膜内皮之间的关系、pIOL后表面与晶状体前表面之间的关系以及瞳孔直径进行动态测量。通过以1.00屈光度(D)的步长从+1.00 D到 -7.00 D增加镜片来刺激生理性调节。

结果

两组在调节时瞳孔直径(P<0.0001,广义线性模型[GLM])以及pIOL前表面与角膜内皮之间的距离均显著减小(P<0.0001,GLM)。两种pIOL的后表面与晶状体前表面之间的距离均无统计学上的显著变化(P = 0.2845,GLM)。两种pIOL在任何测量中均无统计学上的显著差异(P>0.05,GLM)。

结论

结果符合亥姆霍兹调节理论,因为观察到虹膜 - 晶状体隔膜向前移动。pIOL与角膜内皮之间的距离以及瞳孔直径减小,而在整个调节检查过程中,两种pIOL与晶状体之间的距离保持恒定。这表明因调节努力导致晶状体与人工晶状体间歇性接触而引发白内障的风险不大。