用于剂量学应用的非晶硅电子射野影像装置校准:基于响应的蒙特卡罗预测方法与现有技术的比较

Amorphous silicon EPID calibration for dosimetric applications: comparison of a method based on Monte Carlo prediction of response with existing techniques.

作者信息

Parent L, Fielding A L, Dance D R, Seco J, Evans P M

机构信息

Joint Department of Physics, The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, Sutton, UK.

出版信息

Phys Med Biol. 2007 Jun 21;52(12):3351-68. doi: 10.1088/0031-9155/52/12/003. Epub 2007 May 16.

Abstract

For EPID dosimetry, the calibration should ensure that all pixels have a similar response to a given irradiation. A calibration method (MC), using an analytical fit of a Monte Carlo simulated flood field EPID image to correct for the flood field image pixel intensity shape, was proposed. It was compared with the standard flood field calibration (FF), with the use of a water slab placed in the beam to flatten the flood field (WS) and with a multiple field calibration where the EPID was irradiated with a fixed 10x10 field for 16 different positions (MF). The EPID was used in its normal configuration (clinical setup) and with an additional 3 mm copper slab (modified setup). Beam asymmetry measured with a diode array was taken into account in MC and WS methods. For both setups, the MC method provided pixel sensitivity values within 3% of those obtained with the MF and WS methods (mean difference<1%, standard deviation<2%). The difference of pixel sensitivity between MC and FF methods was up to 12.2% (clinical setup) and 11.8% (modified setup). MC calibration provided images of open fields (5x5 to 20x20 cm2) and IMRT fields to within 3% of that obtained with WS and MF calibrations while differences with images calibrated with the FF method for fields larger than 10x10 cm2 were up to 8%. MC, WS and MF methods all provided a major improvement on the FF method. Advantages and drawbacks of each method were reviewed.

摘要

对于电子射野影像装置(EPID)剂量测定,校准应确保所有像素对给定照射具有相似的响应。提出了一种校准方法(MC),即使用蒙特卡罗模拟的射野均匀照射EPID图像的解析拟合来校正射野图像像素强度形状。将其与标准射野校准(FF)、使用置于射束中的水模体使射野变平的校准方法(WS)以及用固定的10×10野对EPID进行16个不同位置照射的多野校准方法(MF)进行比较。EPID采用其正常配置(临床设置)以及额外加一块3毫米铜板的配置(改进设置)。MC和WS方法考虑了用二极管阵列测量的射束不对称性。对于两种设置,MC方法提供的像素灵敏度值与MF和WS方法获得的值相差在3%以内(平均差异<1%,标准差<2%)。MC与FF方法之间的像素灵敏度差异高达12.2%(临床设置)和11.8%(改进设置)。MC校准提供的开放野(5×5至20×20平方厘米)和调强放疗野的图像与WS和MF校准获得的图像相差在3%以内,而对于大于10×10平方厘米的野,与用FF方法校准的图像的差异高达8%。MC、WS和MF方法都比FF方法有显著改进。文中回顾了每种方法的优缺点。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验