An exposimetry system using tissue-mimicking liquid.

作者信息

Stiles Timothy A, Madsen Ernest L, Frank Gary R

机构信息

Department of Medical Physics, University of Wisconsin-Madison, Madison, WI 53706, USA.

出版信息

Ultrasound Med Biol. 2008 Jan;34(1):123-36. doi: 10.1016/j.ultrasmedbio.2007.06.023. Epub 2007 Aug 27.

Abstract

Acoustic output measurements of diagnostic ultrasound scanners are currently performed in water and derated to approximate in situ values. The derating scheme ignores nonlinear propagation of sound waves and has been shown in previous numerical and experimental studies to tend to underestimate relevant pressure and intensity values in tissue mimicking media. This work describes an alternative method, which uses a tissue-mimicking liquid with attenuation coefficient slope of 0.3 dB/cm/MHz, speed of sound of 1,540 m/s and nonlinearity parameter B/A of 7.5. The acoustic properties of this liquid are stable for at least 2 y after production. Initial results using a single M-mode configuration are presented. These results confirm that derating can significantly underestimate the pulse intensity integral and peak rarefactional pressure.

摘要

目前,诊断超声扫描仪的声输出测量是在水中进行的,并进行降额处理以近似原位值。降额方案忽略了声波的非线性传播,并且在先前的数值和实验研究中已表明,该方案往往会低估组织模拟介质中的相关压力和强度值。这项工作描述了一种替代方法,该方法使用一种组织模拟液体,其衰减系数斜率为0.3 dB/cm/MHz,声速为1540 m/s,非线性参数B/A为7.5。这种液体的声学特性在生产后至少2年内是稳定的。给出了使用单一M模式配置的初步结果。这些结果证实,降额处理会显著低估脉冲强度积分和峰值稀疏压力。