Weismüller P, Mayer U, Richter P, Heieck F, Kochs M, Hombach V
Department of Cardiology, Angiology and Pneumology, University of Ulm, Germany.
Eur Heart J. 1991 Nov;12(11):1234-9. doi: 10.1093/eurheartj/12.11.1234.
This study was set up to discover how a subendocardial application of ethanol administered via a catheter would affect an animal model. A 7 F bipolar catheter with a lumen, through which a 2 mm needle was inserted, was placed in the left ventricle of 11 pigs. Altogether, 33 subendocardial injections into the left ventricular myocardium were performed under fluoroscopic control using a mixture of 0.5-1.5 ml ethanol and 0.5-1 ml iopamidol as contrast medium. The mixtures were injected into the apical, lateral and septal walls of the left ventricle. After 25 days, the hearts were removed and the lesions examined pathologically. The calculated volume of the lesions was about 60 mm3, the area in projection to the endocardium about 35 mm2, the depth relative to the endocardium about 1.5 mm and the maximum diameter about 8 mm. Perforation of the myocardial wall by pericardial injection occurred twice without further complications. Subendocardial application of concentrated ethanol by catheter caused a controlled local necrosis. This technique may become a new approach with which to treat ventricular tachycardia by chemical ablation.
本研究旨在探究通过导管在心内膜下应用乙醇会如何影响动物模型。将一根带有管腔的7F双极导管置于11头猪的左心室,通过该管腔插入一根2mm的针头。在透视控制下,使用0.5 - 1.5ml乙醇与0.5 - 1ml碘帕醇作为造影剂的混合物,共对左心室心肌进行了33次心内膜下注射。混合物被注入左心室的心尖、侧壁和间隔壁。25天后,取出心脏并进行病理检查。计算得出的病变体积约为60mm³,心内膜投影面积约为35mm²,相对于心内膜的深度约为1.5mm,最大直径约为8mm。心包注射导致心肌壁穿孔两次,但未出现进一步并发症。通过导管在心内膜下应用浓缩乙醇可导致可控的局部坏死。该技术可能成为一种通过化学消融治疗室性心动过速的新方法。