Li Haitao, Leung Christopher Kai Shun, Wong Lee, Cheung Carol Yim Lui, Pang Chi Pui, Weinreb Robert Neal, Lam Dennis Shun Chiu
Department of Ophthalmology and Visual Sciences, Chinese University of Hong Kong, Hong Kong, China.
Ophthalmology. 2008 May;115(5):796-801.e2. doi: 10.1016/j.ophtha.2007.07.006. Epub 2007 Oct 4.
To evaluate the repeatability and reproducibility of central corneal thickness (CCT) measurements obtained by 2 anterior segment optical coherence tomography (OCT) imaging systems and to examine their agreements with ultrasound pachymetry.
Observational cross-sectional study.
Fifty eyes from 50 healthy normal subjects were recruited.
In one randomly selected eye in each subject, CCT was measured by slit-lamp OCT (SLOCT), Visante OCT, and ultrasound pachymetry. For anterior segment OCT measurements, both automatic and manual CCTs were obtained. Twenty-five of the 50 subjects were invited for 2 more visits within a week to evaluate repeatability and reproducibility of CCT measurement.
Central corneal thickness measurement obtained by the 3 methods and their agreements. Intrasession and intersession within-subject standard deviation (S(w)), precision (1.96xS(w)), coefficient of variation (CV(w)) (100xS(w)/overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility.
Good repeatability and reproducibility were found for both automatic and manual CCT measurements obtained by SLOCT and Visante OCT. For intrasession repeatability, CV(w) and ICC values ranged between 0.9% and 1.2% and 0.96 and 0.98, respectively. For intersession reproducibility, the respective CV(w) and ICC values ranged between 1.2% and 1.4% and 0.94 and 0.96. Although no significant difference was found between automatic/manual SLOCT measurements and ultrasound pachymetry, automatic Visante OCT CCT (535.7+/-30.2 microm) was significantly less than CCT with ultrasound pachymetry (550.3+/-31.14 microm) (P<0.001). In contrast, manual Visante OCT measurement (558.8+/-32.8 microm) was slightly higher than ultrasound pachymetry (P<0.001). Nevertheless, SLOCT and Visante OCT measurement of CCT had 95% limits of agreement comparable to that of ultrasound pachymetry. The best agreement was observed in the manual SLOCT measurement (95% limits of agreement between -15.5 and 11.7 microm).
Both SLOCT and Visante OCT automatic and manual CCT measurements were reliable and showed comparable agreement with ultrasound pachymetry. Although the 2 anterior segment OCT imaging systems have similar design and working principles, clinicians should be aware of the differences in CCT measurement between the 2 anterior segment OCTs.
评估两种眼前节光学相干断层扫描(OCT)成像系统所获得的中央角膜厚度(CCT)测量值的可重复性和再现性,并检验其与超声角膜测厚法的一致性。
观察性横断面研究。
招募了50名健康正常受试者的50只眼睛。
在每个受试者随机选择的一只眼睛中,通过裂隙灯OCT(SLOCT)、Visante OCT和超声角膜测厚法测量CCT。对于眼前节OCT测量,同时获取自动和手动测量的CCT值。50名受试者中的25名在一周内被邀请进行另外2次检查,以评估CCT测量的可重复性和再现性。
通过3种方法获得的中央角膜厚度测量值及其一致性。计算测量期间内和测量期间间受试者内标准差(S(w))、精密度(1.96×S(w))、变异系数(CV(w))(100×S(w)/总体均值)和组内相关系数(ICC),以评估可重复性和再现性。
SLOCT和Visante OCT获得的自动和手动CCT测量值均具有良好的可重复性和再现性。对于测量期间内的可重复性,CV(w)值和ICC值分别在0.9%至1.2%以及0.96至0.98之间。对于测量期间间的再现性,相应的CV(w)值和ICC值分别在1.2%至1.4%以及0.94至0.96之间。虽然自动/手动SLOCT测量值与超声角膜测厚法之间未发现显著差异,但自动Visante OCT测量的CCT(535.7±30.2微米)显著低于超声角膜测厚法测量的CCT(550.3±31.14微米)(P<0.001)。相比之下,手动Visante OCT测量值(558.8±32.8微米)略高于超声角膜测厚法(P<0.001)。然而,SLOCT和Visante OCT测量的CCT具有与超声角膜测厚法相当的95%一致性界限。在手动SLOCT测量中观察到最佳一致性(95%一致性界限在-15.5至11.7微米之间)。
SLOCT和Visante OCT的自动和手动CCT测量均可靠,且与超声角膜测厚法具有相当的一致性。尽管这两种眼前节OCT成像系统具有相似的设计和工作原理,但临床医生应注意这两种眼前节OCT在CCT测量上的差异。