Tearney G J, Webb R H, Bouma B E
Wellman Laboratories of Photomedicine, Massachusetts General Hospital, 50 Blossom Street, BAR 703, Boston, Massachusetts 02114, USA.
Opt Lett. 1998 Aug 1;23(15):1152-4. doi: 10.1364/ol.23.001152.
An endoscope-compatible, submicrometer-resolution scanning confocal microscopy imaging system is presented. This approach, spectrally encoded confocal microscopy (SECM), uses a quasi-monochromatic light source and a transmission diffraction grating to detect the reflectivity simultaneously at multiple points along a transverse line within the sample. Since this method does not require fast spatial scanning within the probe, the equipment can be miniaturized and incorporated into a catheter or endoscope. Confocal images of an electron microscope grid were acquired with SECM to demonstrate the feasibility of this technique.
本文介绍了一种与内窥镜兼容的亚微米分辨率扫描共聚焦显微镜成像系统。这种方法,即光谱编码共聚焦显微镜(SECM),使用准单色光源和透射衍射光栅在样品内沿横向线的多个点同时检测反射率。由于该方法不需要探头内的快速空间扫描,因此设备可以小型化并集成到导管或内窥镜中。用SECM采集了电子显微镜网格的共聚焦图像,以证明该技术的可行性。